Recall Register

Device recall: Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit…

Class II FDA enforcement event 78583. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Implant Direct Sybron Manufacturing, LLC started this device recall on 15 Nov 2017. The firm address in the FDA record is in Westlake Village, CA. FDA put the recall in Class II. FDA event 78583 has 1 product record. The first FDA enforcement report date is 27 Dec 2017. The record gives the status Terminated, with the termination date 9 Oct 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.

Recalling firm

Implant Direct Sybron Manufacturing, LLC, Westlake Village, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 78583: product records
Recall numberProductQuantityClassStatusReport date
Z-0283-2018Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The…560 units (worldwide)Class IITerminated27 Dec 2017

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2017
1,091
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
30

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 78583. https://recallregister.com/fda/ii/78583/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.
What quantity of product is in FDA event 78583?
The FDA record gives this quantity: 560 units (worldwide)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2017 have the same class?
93% of the 1,091 device events that started in 2017 are in Class II.