Recall Register

Device recall: Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

Class II FDA enforcement event 78130. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Handicare AB started this device recall on 28 Jul 2016. The firm address in the FDA record is in Lulea. FDA put the recall in Class II. FDA event 78130 has 1 product record. The first FDA enforcement report date is 18 Oct 2017. The record gives the status Terminated, with the termination date 17 Dec 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

OH, PA, MD

Recalling firm

Handicare AB, Lulea, Sweden.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 78130: product records
Recall numberProductQuantityClassStatusReport date
Z-0014-2018Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 606000099 unitsClass IITerminated18 Oct 2017

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
3

Other events of the firm

Other FDA enforcement events of Handicare AB
InitiatedProduct (first record)TypeClassRecords
21 Aug 2020SwiftHook sold as the following: a. Individually as SwiftHook b. Included with Vega505EE mobile lifts Product Usage: TheDeviceClass II1
15 Jun 2016EVA Floor Lifts, Model Numbers: 400, 450, 600DeviceClass II1

Handicare AB: all 3 events

Official channels

Cite this page

Recall Register. FDA enforcement event 78130. https://recallregister.com/fda/ii/78130/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: OH, PA, MD
What quantity of product is in FDA event 78130?
The FDA record gives this quantity: 9 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.