Device recall: Alaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP…
Class II FDA enforcement event 77987. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The recalling firm has received reports of increased or decreased flows that have occurred in certain pumps.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
CareFusion 303, Inc. started this device recall on 9 Aug 2017. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class II. FDA event 77987 has 1 product record. The first FDA enforcement report date is 25 Oct 2017. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - USA (nationwide) and to the countries of : Canada, Australia, UAE, Kuwait, Saudi Arabia, South Africa
Recalling firm
CareFusion 303, Inc., San Diego, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0026-2018 | Alaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N) 10942012 between November 2011 and March 2012; Alaris Pump module Bezel Kit Assembly (P/N) 10964559) shipped between November 2011 and March 2012. The Alaris Pump module… | 31,622 units (28,224 in US) | Class II | Ongoing | 25 Oct 2017 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2017
- 1,091
- Share of these events in Class II
- 93%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 99
Other events of the firm
Official channels
- FDA enforcement report Z-0026-2018The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 77987. https://recallregister.com/fda/ii/77987/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - USA (nationwide) and to the countries of : Canada, Australia, UAE, Kuwait, Saudi Arabia, South Africa
- What quantity of product is in FDA event 77987?
- The FDA record gives this quantity:
31,622 units (28,224 in US)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2017 have the same class?
- 93% of the 1,091 device events that started in 2017 are in Class II.