Device recall: iVue with Normative Database, Device Listing Number D185549. Product Usage: The iVue (K121739) is a non-contact, high…
Class II FDA enforcement event 77776. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Optovue, Inc. started this device recall on 27 Jul 2017. The firm address in the FDA record is in Fremont, CA. FDA put the recall in Class II. FDA event 77776 has 3 product records. The first FDA enforcement report date is 6 Sep 2017. The record gives the status Terminated, with the termination date 12 Apr 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA,…
Recalling firm
Optovue, Inc., Fremont, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0008-2018 | iVue with Normative Database, Device Listing Number D185549. Product Usage: The iVue (K121739) is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three dimensional imaging and measurement of anterior and posterior ocular structures, including… | 326 devices | Class II | Terminated | 6 Sep 2017 |
| Z-0009-2018 | iFusion, Device Listing Number D203271. Product Usage: The iCam is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. The iCam takes digital images of the… | 326 devices | Class II | Terminated | 6 Sep 2017 |
| Z-0010-2018 | iVue 500 with iScan, D216606. Product non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC),… | 326 devices | Class II | Terminated | 6 Sep 2017 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2017
- 1,091
- Share of these events in Class II
- 93%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 4
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 16 Mar 2016 | iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of | Device | Class II | 1 |
| 14 May 2013 | Optovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging. | Device | Class III | 1 |
| 7 Sep 2012 | Reference materials "RETINAL OCT Analysis and Interpretation Method" shipped with iVue System (K091404) and the "Retina | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0008-2018The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 77776. https://recallregister.com/fda/ii/77776/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA,…
- What quantity of product is in FDA event 77776?
- The event has 3 product records. The first record gives this quantity:
326 devices
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2017 have the same class?