Device recall: Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for…

Class II FDA enforcement event 77631. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Angiodynamics, Inc. started this device recall on 5 Jun 2017. The firm address in the FDA record is in Queensbury, NY. FDA put the recall in Class II. FDA event 77631 has 6 product records. The first FDA enforcement report date is 14 Feb 2018. The record gives the status Terminated, with the termination date 11 Sep 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide

Recalling firm

Angiodynamics, Inc., Queensbury, NY.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 77631: product records
Recall numberProductQuantityClassStatusReport date
Z-0526-2018Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018
Z-0527-2018Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018
Z-0528-2018Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602, Model # (UPN) H7879006020, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018
Z-0529-2018Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # (UPN) H787900600US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018
Z-0530-2018Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018
Z-0531-2018Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602-US, Model # (UPN) H787900602US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.14,353 units in totalClass IITerminated14 Feb 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2017
1,091
Share of these events in Class II
93%
Product records in this event
6Median of the firm: 2
Events of the firm on the register
24

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 77631. https://recallregister.com/fda/ii/77631/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide
What quantity of product is in FDA event 77631?
The event has 6 product records. The first record gives this quantity: 14,353 units in total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2017 have the same class?
93% of the 1,091 device events that started in 2017 are in Class II.