Recall Register

Device recall: EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI…

Class II FDA enforcement event 76498. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Becton Dickinson & Co. started this device recall on 25 Aug 2014. The firm address in the FDA record is in Sparks, MD. FDA put the recall in Class II. FDA event 76498 has 1 product record. The first FDA enforcement report date is 15 Mar 2017. The record gives the status Terminated, with the termination date 8 Mar 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors

Recalling firm

Becton Dickinson & Co., Sparks, MD.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 76498: product records
Recall numberProductQuantityClassStatusReport date
Z-1414-2017EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data…Qty Distr. to Field - 115Class IITerminated15 Mar 2017

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
77

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 76498. https://recallregister.com/fda/ii/76498/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: 15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
What quantity of product is in FDA event 76498?
The FDA record gives this quantity: Qty Distr. to Field - 115
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.