Device recall: BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 3.4MM(D) X 2MM(H) An endosseous dental implant abutment is a…

Class II FDA enforcement event 75761. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

A limited number of product pouches may not have been sealed during packaging.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Biomet 3i, LLC started this device recall on 14 Sep 2016. The firm address in the FDA record is in Palm Beach Gardens, FL. FDA put the recall in Class II. FDA event 75761 has 36 product records. The first FDA enforcement report date is 28 Dec 2016. The record gives the status Terminated, with the termination date 26 Jul 2019. 5 of the 1,232 device events that started in 2016 have more product records. The median device event of 2016 has 1 product record. The openFDA file has 1,232 device events that started in 2016 and 1,153 that started in 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution to Austria, Belgium, Switzerland, Germany, Spain, France, Great Britain, Ireland, Italy, Lithuania, Portugal and Sweden

Recalling firm

Biomet 3i, LLC, Palm Beach Gardens, FL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

36 product records in the event.

FDA event 75761: product records
Recall numberProductQuantityClassStatusReport date
Z-0828-2017BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 3.4MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation127Class IITerminated28 Dec 2016
Z-0829-2017BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 3.4MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation53Class IITerminated28 Dec 2016
Z-0830-2017BIOMET 3i CERTAIN LOW PROFILE 30¿ ABUTMENT 3.4MM(D) X 5MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation27Class IITerminated28 Dec 2016
Z-0831-2017BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 4.1MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation735Class IITerminated28 Dec 2016
Z-0832-2017BIOMET 3i CERTAIN LOW PROFILE 30¿ ABUTMENT 4.1MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation431Class IITerminated28 Dec 2016
Z-0833-2017BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 4.1MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation150Class IITerminated28 Dec 2016
Z-0834-2017BIOMET 3i CERTAIN LOW PROFILE 30¿ ABUTMENT 4.1MM(D) X 5MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation228Class IITerminated28 Dec 2016
Z-0835-2017BIOMET 3i CERTAIN LOW PROFILE ABUTMENT 3.4MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation50Class IITerminated28 Dec 2016
Z-0836-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation101Class IITerminated28 Dec 2016
Z-0837-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation64Class IITerminated28 Dec 2016
Z-0838-2017BIOMET 3i CERTAIN LOW PROFILE ABUTMENT 4.1MM(D) X 1MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation123Class IITerminated28 Dec 2016
Z-0839-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 1MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation450Class IITerminated28 Dec 2016
Z-0840-2017BIOMET 3i CERTAIN LOW PROFILE ABUTMENT 4.1MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation108Class IITerminated28 Dec 2016
Z-0841-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation525Class IITerminated28 Dec 2016
Z-0842-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation46Class IITerminated28 Dec 2016
Z-0843-2017BIOMET 3i CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation108Class IITerminated28 Dec 2016
Z-0844-2017BIOMET 3i LOW PROFILE 30¿ ABUTMENT 3.4MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation20Class IITerminated28 Dec 2016
Z-0845-2017BIOMET 3i LOW PROFILE 30¿ ABUTMENT 3.4MM(D) X 5MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation4Class IITerminated28 Dec 2016
Z-0846-2017BIOMET 3i LOW PROFILE 17¿ ABUTMENT 4.1MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation246Class IITerminated28 Dec 2016
Z-0847-2017BIOMET 3i LOW PROFILE 30¿ ABUTMENT 4.1MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation175Class IITerminated28 Dec 2016
Z-0848-2017BIOMET 3i LOW PROFILE 17¿ ABUTMENT 4.1MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation216Class IITerminated28 Dec 2016
Z-0849-2017BIOMET 3i LOW PROFILE ABUTMENT 4.1MM(D) X 1MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation160Class IITerminated28 Dec 2016
Z-0850-2017BIOMET 3i LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 1MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation204Class IITerminated28 Dec 2016
Z-0851-2017BIOMET 3i LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 2MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation137Class IITerminated28 Dec 2016
Z-0852-2017BIOMET 3i LOW PROFILE ABUTMENT 4.1MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation50Class IITerminated28 Dec 2016
Z-0853-2017BIOMET 3i LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 3MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation191Class IITerminated28 Dec 2016
Z-0854-2017BIOMET 3i LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 4MM(H) An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation124Class IITerminated28 Dec 2016
Z-0855-2017BIOMET 3i LOW PROFILE ABUTMENT HEALING CAP An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation391Class IITerminated28 Dec 2016
Z-0856-2017BIOMET 3i LOW PROFILE ABUTMENT (HEXED) PICK-UP IMPRESSION COPING An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation28Class IITerminated28 Dec 2016
Z-0857-2017BIOMET 3i LOW PROFILE ABUTMENT (NON-HEXED) PICK-UP IMPRESSION COPING An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation458Class IITerminated28 Dec 2016
Z-0858-2017BIOMET 3i LOW PROFILE ABUTMENT NON-HEXED PREFORMANCE TEMPORARY CYLINDER An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation270Class IITerminated28 Dec 2016
Z-0859-2017BIOMET 3i LOW PROFILE ABUTMENT QUICKBRIDGE An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation60Class IITerminated28 Dec 2016
Z-0860-2017BIOMET 3i LOW PROFILE ABUTMENT HEXED TEMPORARY CYLINDER An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation108Class IITerminated28 Dec 2016
Z-0861-2017BIOMET 3i LOW PROFILE ABUTMENT NON-HEXED TEMPORARY CYLINDER An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation1,371Class IITerminated28 Dec 2016
Z-0862-2017BIOMET 3i LOW PROFILE ABUTMENT (NON-HEXED) TRANSFER IMPRESSION COPING An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation191Class IITerminated28 Dec 2016
Z-0863-2017BIOMET 3i LOW PROFILE ABUTMENT TITANIUM RETAINING SCREW An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation1,633Class IITerminated28 Dec 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
36Median of the firm: 2
Events of the firm on the register
24

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 75761. https://recallregister.com/fda/ii/75761/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution to Austria, Belgium, Switzerland, Germany, Spain, France, Great Britain, Ireland, Italy, Lithuania, Portugal and Sweden
What quantity of product is in FDA event 75761?
The event has 36 product records. The first record gives this quantity: 127
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.