Device recall: PET Infusion Pump
Class II FDA enforcement event 75387. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Bayer Healthcare started this device recall on 30 Sep 2016. The firm address in the FDA record is in Indianola, PA. FDA put the recall in Class II. FDA event 75387 has 1 product record. The first FDA enforcement report date is 23 Nov 2016. The record gives the status Terminated, with the termination date 24 Mar 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Distributed Nationwide and internationally to : Taiwan, Ukraine, Turkey, Thailand, Slovania, Singapore, Sweden, Saudi Arabia, Serbia, Romania, Qatar, Portugal, Poland, Panama, Oman, Norway, The…
Recalling firm
Bayer Healthcare, Indianola, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0400-2017 | PET Infusion Pump | 165,127 (79,714 units in U.S.) | Class II | Terminated | 23 Nov 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 6
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 7 Nov 2016 | MEDRAD Intego PET Infusion System - Pump Intended to deliver accurate doses of Fluoredeoxyglucose or F-Sodium Fluoride a | Device | Class II | 1 |
| 2 Aug 2016 | Certegra Workstation used in conjunction with Medrad Stellant CT Injection System Catalog Number # CWKS DISPLAY Serial N | Device | Class II | 1 |
| 9 Jun 2016 | Angiographic Injector and Syringe, MEDRAD MRXperion MR Injection System, MEDRAD MRXperion Sterile Disposable MRI Kit Usa | Device | Class II | 1 |
| 30 Nov 2015 | Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nucl | Device | Class II | 1 |
| 4 Aug 2015 | Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be use | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0400-2017The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 75387. https://recallregister.com/fda/ii/75387/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Distributed Nationwide and internationally to : Taiwan, Ukraine, Turkey, Thailand, Slovania, Singapore, Sweden, Saudi Arabia, Serbia, Romania, Qatar, Portugal, Poland, Panama, Oman, Norway, The…
- What quantity of product is in FDA event 75387?
- The FDA record gives this quantity:
165,127 (79,714 units in U.S.)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?