Device recall: Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190
Class II FDA enforcement event 75366. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
.Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Roche Diagnostics Operations, Inc. started this device recall on 4 Oct 2016. The firm address in the FDA record is in Indianapolis, IN. FDA put the recall in Class II. FDA event 75366 has 2 product records. The first FDA enforcement report date is 23 Nov 2016. The record gives the status Terminated, with the termination date 5 Jun 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Nationwide.
Recalling firm
Roche Diagnostics Operations, Inc., Indianapolis, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0395-2017 | Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190 | 4976 | Class II | Terminated | 23 Nov 2016 |
| Z-0396-2017 | COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160 | 4976 | Class II | Terminated | 23 Nov 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 69
Other events of the firm
Official channels
- FDA enforcement report Z-0395-2017The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 75366. https://recallregister.com/fda/ii/75366/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Nationwide.
- What quantity of product is in FDA event 75366?
- The event has 2 product records. The first record gives this quantity:
4976
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?
- 91% of the 1,232 device events that started in 2016 are in Class II.