Device recall: EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter
Class II FDA enforcement event 75046. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
EKOS Corporation started this device recall on 9 Aug 2016. The firm address in the FDA record is in Bothell, WA. FDA put the recall in Class II. FDA event 75046 has 1 product record. The first FDA enforcement report date is 28 Sep 2016. The record gives the status Terminated, with the termination date 5 Oct 2016.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Three international consignees - Germany, New Zealand and Taiwan
Recalling firm
EKOS Corporation, Bothell, WA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2854-2016 | EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter | 10 units | Class II | Terminated | 28 Sep 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 6
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 22 Aug 2016 | The EKOS cart has the catalog number 700-60101. The EKOS cart is a custom infusion stand/cart used to hold and transport | Device | Class II | 1 |
| 21 Jan 2016 | EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endo | Device | Class II | 1 |
| 24 Jun 2015 | The EKOS EkoSonic Control Unit is intended exclusively for use with the EkoSonic Endovascular Device and the EkoSonic MA | Device | Class II | 1 |
| 2 Oct 2013 | The EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per transducer) | Device | Class II | 1 |
| 24 May 2012 | EkoSonic MACH4 Endovascular Device (Catalog #500-55106, 500-55112; 500-55118; 500-55124; 500-55130; 500-55140; 500-55150 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-2854-2016The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 75046. https://recallregister.com/fda/ii/75046/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Three international consignees - Germany, New Zealand and Taiwan
- What quantity of product is in FDA event 75046?
- The FDA record gives this quantity:
10 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?