Device recall: NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011…
Class II FDA enforcement event 74975. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Aribex Inc started this device recall on 18 Jul 2016. The firm address in the FDA record is in Charlotte, NC. FDA put the recall in Class II. FDA event 74975 has 2 product records. The first FDA enforcement report date is 14 Sep 2016. The record gives the status Terminated, with the termination date 26 Jul 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI,…
Recalling firm
Aribex Inc, Charlotte, NC.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2716-2016 | NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019 | 16,871 | Class II | Terminated | 14 Sep 2016 |
| Z-2717-2016 | NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GBL, Part Number 08500025 and Part… | 16,871 | Class II | Terminated | 14 Sep 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 2
- Events of the firm on the register
- 1
Official channels
- FDA enforcement report Z-2716-2016The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 74975. https://recallregister.com/fda/ii/74975/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI,…
- What quantity of product is in FDA event 74975?
- The event has 2 product records. The first record gives this quantity:
16,871
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?
- 91% of the 1,232 device events that started in 2016 are in Class II.