Device recall: Transpac IV Monitoring Kit with SafeSet 84inch Arterial Pressure Tubing, Reservoir, Squeeze Flush and 2 Needleless…

Class II FDA enforcement event 74966. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

ICU Medical, Inc. started this device recall on 10 Aug 2016. The firm address in the FDA record is in San Clemente, CA. FDA put the recall in Class II. FDA event 74966 has 24 product records. The first FDA enforcement report date is 21 Sep 2016. The record gives the status Terminated, with the termination date 10 Jul 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide Distribution

Recalling firm

ICU Medical, Inc., San Clemente, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

24 product records in the event.

FDA event 74966: product records
Recall numberProductQuantityClassStatusReport date
Z-2805-2016Transpac IV Monitoring Kit with SafeSet 84inch Arterial Pressure Tubing, Reservoir, Squeeze Flush and 2 Needleless Valves, Item No. 011-42646-66 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by…420 unitsClass IITerminated21 Sep 2016
Z-2806-2016Transpac IV Trifurcated Kit With Safeset Reservoir And 2 Blood Sampling Ports, 84 inch Tubing, 3 3 ml Intraflo Flushes, 3 Disposable Transducers, Macrodrip, Item No. 42500-04 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter…250 unitsClass IITerminated21 Sep 2016
Z-2807-2016Transpac IV Bifurcated Kit with SafeSet Reservoir, 03 ml Squeeze Flush and Needleless Valve, Item No. 42500-66 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into…10 unitsClass IITerminated21 Sep 2016
Z-2808-2016Transpac IV Monitoring Kit with 2 SafeSet Blood Sampling Ports, 60" TUBING, Disposable Transducer, 3 ML Intraflo Flush, Macrodrip (Pole Mount), Item No. 42619-08 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure…60 unitsClass IITerminated21 Sep 2016
Z-2809-2016Transpac IV Monitoring Kit with 2 SafeSet Reservoir and 2 Blood Sampling Ports, 84" Tubing, Disposable Transducer, 3 mL Squeeze Flush, Macrodrip (Pole Mount), Item No. 42646-06 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter…4180 unitsClass IITerminated21 Sep 2016
Z-2810-2016Transpac IV Monitoring Kit with 2 SafeSet 84" Arterial Pressure Tubing, Reservoir, 03 mL Squeeze Flush and 2 Needleless Valves, Item No. 42646-66 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by…1420 unitsClass IITerminated21 Sep 2016
Z-2811-2016Transpac IV Monitoring Kit with SafeSet Reservoir and Blood Sampling Port, 24" Tubing, Disposable Transducer, 03 mL Squeeze Flush, Macrodrip (Patient Mount), Item No. 42648-06 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a…980 unitsClass IITerminated21 Sep 2016
Z-2812-2016Transpac IV Trifurcated Monitoring Kit with SafeSet Reservoir and 2 Blood Sampling Ports, 84", 3.3mL Squeeze Flushes, 3 Disposable Transducers, Macrodrip (Pole Mount), Item No. 42650-06 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces…760 unitsClass IITerminated21 Sep 2016
Z-2813-2016Transpac IV Trifurcated Monitoring Kit with 84" SafeSet Reservoir, 03 ml Squeeze Flush and Needleless Valves, Item No. 42650-66 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting…90 unitsClass IITerminated21 Sep 2016
Z-2814-2016Single Monitoring Kit with 30mL Flush Device, Item No. 46073-38 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a…20 unitsClass IITerminated21 Sep 2016
Z-2815-2016Monitoring Kit with 30mL Flush Device, Item No. 46076-15 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a…10 unitsClass IITerminated21 Sep 2016
Z-2816-2016Triple Line SafeSet Kit with 03mL Flush Device, Item No. 46076-27 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into…150 unitsClass IITerminated21 Sep 2016
Z-2817-2016Double Kit with SafeSet and 03mL Flush Device, Item No. 46080-37 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into…110 unitsClass IITerminated21 Sep 2016
Z-2818-2016Monitoring Kit with 03mL Flush Device, Item No. 46080-57 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a…70 unitsClass IITerminated21 Sep 2016
Z-2819-2016SafeSet Kit with 30 mL Flush Device, Item No. 46083-42 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a…125 unitsClass IITerminated21 Sep 2016
Z-2820-2016Bifurcated SafeSet Kit with 03mL Flush Device, Item No. 46091-75 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into…20 unitsClass IITerminated21 Sep 2016
Z-2821-2016Pediatric Kit with 30mL Flush Device, Item No. 46099-10 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a…20 unitsClass IITerminated21 Sep 2016
Z-2822-20165 Port OFF Manifold with Transpac IV, Item No. 46100-29 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible…10 unitsClass IITerminated21 Sep 2016
Z-2823-2016Monitoring Kit with TP4, 30 ml Squeeze Flush and Needleless Valve, Item No. 46103-22 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be…20 unitsClass IITerminated21 Sep 2016
Z-2824-2016Transpac IV Monitoring Kit with SafeSet 84" Arterial Pressure Tubing, Reservoir, Squeeze Flush and Needleless Valve, Item No. 46104-53 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting…100 unitsClass IITerminated21 Sep 2016
Z-2825-2016TP4 Kit with 10cc SafeSet Reservoir, Needleless Valve, Filter Millex and Admin Set, Item No. 46104-65 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical…140 unitsClass IITerminated21 Sep 2016
Z-2826-2016Transpac IV Monitoring Kit with 03 ml Squeeze Flush, SafeSet Reservoir, 2 Needleless Valves, Mounts and Bracket, Item Nos. 46107-35, 46110-07 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting…70 unitsClass IITerminated21 Sep 2016
Z-2827-2016Transpac IV Trifurcated Monitoring Kit with SafeSet Reservoir, Port and 03 ml Squeeze Flush Devices, Item No. 46110-84 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure…150 unitsClass IITerminated21 Sep 2016
Z-2828-2016Monitoring Kit with TP4, 30 ml Squeeze Flush Device, 10 cc Contamination Syringe and Needleless Valve, Item No. 46112-65 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in…20 unitsClass IITerminated21 Sep 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
24Median of the firm: 2
Events of the firm on the register
75

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 74966. https://recallregister.com/fda/ii/74966/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide Distribution
What quantity of product is in FDA event 74966?
The event has 24 product records. The first record gives this quantity: 420 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.