Recall Register

Device recall: Spirit Plus Bed, A-C Powered Hospital Bed

Class II FDA enforcement event 74560. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

CHG Hospital Beds Inc started this device recall on 28 Jun 2016. The firm address in the FDA record is in London. FDA put the recall in Class II. FDA event 74560 has 2 product records. The first FDA enforcement report date is 10 Aug 2016. The record gives the status Terminated, with the termination date 8 Jun 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution: US (nationwide) including states of: AK , AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI,…

Recalling firm

CHG Hospital Beds Inc, London, Canada.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 74560: product records
Recall numberProductQuantityClassStatusReport date
Z-2331-2016Spirit Plus Bed, A-C Powered Hospital Bed4,308 units totalClass IITerminated10 Aug 2016
Z-2332-2016Spirit Select Bed, A-C Powered Hospital Bed4,308 units totalClass IITerminated10 Aug 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of CHG Hospital Beds Inc
InitiatedProduct (first record)TypeClassRecords
18 Feb 2015The Spirit TM Select bed is a Med-Surg bed intended to support and transport low to moderate acuity patients in the medi Event 73422DeviceClass II1

Official channels

Cite this page

Recall Register. FDA enforcement event 74560. https://recallregister.com/fda/ii/74560/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution: US (nationwide) including states of: AK , AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI,…
What quantity of product is in FDA event 74560?
The event has 2 product records. The first record gives this quantity: 4,308 units total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.