Device recall: Richard Allan Scientific 10% Neutral Buffered Formalin, Part Number: 53901 and 561201 Item 53901 90ml bottle, 125 per…
Class II FDA enforcement event 74557. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
potential contamination with potassium hydroxide may have an impact when performing immunohistochemistry testing. It is unknown whether the contamination will affect tissue antigenicity. If it were to affect tissue antigenicity, the outcome would be risk of false positive or false negative test results which may lead to serious injury. There is the possibility the patient may need to be…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Richard-Allan Scientific Company started this device recall on 13 Jun 2016. The firm address in the FDA record is in Kalamazoo, MI. FDA put the recall in Class II. FDA event 74557 has 1 product record. The first FDA enforcement report date is 17 Aug 2016. The record gives the status Terminated, with the termination date 8 Nov 2016.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Distributed in the states of FL, PA, NY, MD, CA, and AZ.
Recalling firm
Richard-Allan Scientific Company, Kalamazoo, MI.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2351-2016 | Richard Allan Scientific 10% Neutral Buffered Formalin, Part Number: 53901 and 561201 Item 53901 90ml bottle, 125 per case; Item 561201 120ml bottle, 75 per case. | 18 cases | Class II | Terminated | 17 Aug 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class II
- 91%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 7
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 4 Jun 2018 | EZ Single Cytofunnel Brown, A78710004 Product Usage: The EZ Single Cytofunnel with Brown filter papter are intended for | Device | Class III | 1 |
| 15 Mar 2017 | Shandon Rapid Chrome Kwik-Diff Kit Part Number: 9990700 Shandon Rapid Chrome Kwik-Diff Reagent #3 Part Number 9990707 Sh | Device | Class II | 1 |
| 16 Feb 2016 | Richard-Allan Scientific Paraffin Type 3 For in vitro diagnostic use. For use as an infiltration and embedding medium in | Device | Class III | 1 |
| 29 Apr 2015 | Mucolytic Agent 1 pint/473 ml; 1 gallon/3.8 l Product Usage: A mucoliquefying preservative designed for use in the prepa | Device | Class II | 2 |
| 19 Feb 2015 | Carton: Richard-Allan Scientific CYTO-STAIN; REF 7501R; Quantity: 1x1 Gallon; Size: 1 Gallon/3.8L; LOT 313833; 11/2016; | Device | Class III | 1 |
| 17 Sep 2014 | Protocol 10% NBF Product Usage: 10% NBF is used to store and fixate tissue prior to grossing and histological examinatio | Device | Class I | 7 |
Official channels
- FDA enforcement report Z-2351-2016The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 74557. https://recallregister.com/fda/ii/74557/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Distributed in the states of FL, PA, NY, MD, CA, and AZ.
- What quantity of product is in FDA event 74557?
- The FDA record gives this quantity:
18 cases
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?
- 91% of the 1,232 device events that started in 2016 are in Class II.