Device recall: Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots are…

Class II FDA enforcement event 74207. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Complaints of “smoke” coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later i

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Stryker Medical Division of Stryker Corporation started this device recall on 5 May 2016. The firm address in the FDA record is in Portage, MI. FDA put the recall in Class II. FDA event 74207 has 1 product record. The first FDA enforcement report date is 22 Jun 2016. The record gives the status Terminated, with the termination date 18 Sep 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distributed US (nationwide) and the countries of Australia, Canada, Hong Kong, Spain, Switzerland, and United Kingdom.

Recalling firm

Stryker Medical Division of Stryker Corporation, Portage, MI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 74207: product records
Recall numberProductQuantityClassStatusReport date
Z-1990-2016Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots are designed to support and transport a maximum weight of 700 lb. (318 kg) in pre-hospital and hospital environments.770Class IITerminated22 Jun 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
37

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 74207. https://recallregister.com/fda/ii/74207/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distributed US (nationwide) and the countries of Australia, Canada, Hong Kong, Spain, Switzerland, and United Kingdom.
What quantity of product is in FDA event 74207?
The FDA record gives this quantity: 770
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?