Recall Register

Food, drugs, devices › Biomet, Inc.

Device recall: Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) Biomet Comprehensive Reverse Shoulder

FDA enforcement event 73173 · Class II · initiated 16 Feb 2016 · status Terminated · terminated 30 May 2017

DeviceVoluntary: Firm initiated1 product recordWarsaw, IN

Biomet, Inc. (Warsaw, IN) initiated this device recall on 16 Feb 2016; FDA lists 1 product record under event 73173, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,232 device enforcement events initiated in 2016 carry the same class.

Status: Terminated, terminated 30 May 2017. Initiation: voluntary: firm initiated. Biomet, Inc. has 84 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Biomet, Inc., Warsaw, IN.

Reason for recall

"Zimmer Biomet Inc. Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) supplied by Kirkstall Precision was made according to the W1.4034 (420 stainless) which does not conform to 420 S29 indicated on the print. All units manufactured to date were manufactured using 420 stainless. Therefore, all lots of PN: 402852 are considered in scope."

Distribution

"Worldwide Distribution - US including CA, NY, VA, IL, IN, SC, NC, AR, MO, ND, AL, PA, MI, WA, OR, LA, FL, GA, OH, KS, KY, TX, MA, AZ, NJ, UT, IO, NM, WV, MS, NE, and MD OUS: Australia, Chile,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1093-2016Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly…1261 unitsClass IITerminated16 Mar 2016

Context

91%of 2016 device events are Class II1,232 events that year
1product record in this eventfirm median: 1
84events on this register for the firm

Other enforcement events for Biomet, Inc.

InitiatedProduct (first record)TypeClassRecords
1 Jul 2026Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade ADeviceClass II5
23 Sep 2025BIOMET ZipTight, Acute AC Joint Implant, Single Ziploop , Model/Catalog Number: 904834; Fixation System for AC Joint RepDeviceClass II1
9 Sep 2025Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080DeviceClass II1
21 Oct 2024JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model number 110027357DeviceClass II1
17 May 2024RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206DeviceClass II1
13 Mar 2024Vanguard Knee System AS Tibial Bearing / DCM ArCom, 12 MM X 75 MM; Model Number 189082. The product is placed into a foiDeviceClass II1
2 Nov 2023Biolox Ceramic Option Head Zimmer 12/14 40mm -3, Item Number 00-8777-040-01DeviceClass II2
9 Oct 2023Compress Device Segmental Anchor Plug, 10 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previDeviceClass II19
6 Feb 2023G7 Acetabular System, Dual Mobility Acetabular Liner, 40 mm, Size D, Model Number 110024462DeviceClass II1
29 Dec 2022(1) Vanguard Knee System PS Open Box Femoral; Right; 60 mm. Intended for knee joint replacement prostheses. Item Number:DeviceClass II2

All 84 events for Biomet, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Zimmer Biomet Inc. Verso Shoulder Forked Retractor (Comprehensive Instrumentation 2 Prong Retractor) supplied by Kirkstall Precision was made according to the W1.4034 (420 stainless) which does not conform to 420 S29 indicated on the print. All units manufactured to date were manufactured using 420 stainless. Therefore, all lots of PN: 402852 are considered in scope."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - US including CA, NY, VA, IL, IN, SC, NC, AR, MO, ND, AL, PA, MI, WA, OR, LA, FL, GA, OH, KS, KY, TX, MA, AZ, NJ, UT, IO, NM, WV, MS, NE, and MD OUS: Australia, Chile,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 30 May 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.