Device recall: Allura Xper FD10, FD10 C, and FD10 F; Model Numbers: 722026 722010 722003 722002 722001 Intended for Cardiovascular…

Class II FDA enforcement event 72599. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Electronics North America Corporation started this device recall on 3 Nov 2015. The firm address in the FDA record is in Andover, MA. FDA put the recall in Class II. FDA event 72599 has 29 product records. The first FDA enforcement report date is 10 Feb 2016. The record gives the status Terminated, with the termination date 15 Sep 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and…

Recalling firm

Philips Electronics North America Corporation, Andover, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

29 product records in the event.

FDA event 72599: product records
Recall numberProductQuantityClassStatusReport date
Z-0702-2016Allura Xper FD10, FD10 C, and FD10 F; Model Numbers: 722026 722010 722003 722002 722001 Intended for Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP).13297 in totalClass IITerminated10 Feb 2016
Z-0703-2016Allura Xper FD10/10; Model Numbers: 722027 722011 722005 Intended for Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP).13297 in totalClass IITerminated10 Feb 2016
Z-0704-2016Allura Xper FD10 OR Table; Model Numbers: 722033 722022 Intended for Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP).13297 in totalClass IITerminated10 Feb 2016
Z-0705-2016Allura Xper FD20/10; Model Numbers: 722029 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. " Cardiac imaging…13297 in totalClass IITerminated10 Feb 2016
Z-0706-2016Allura Xper FD20/15; Model Numbers: 722058 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. " Cardiac imaging…13297 in totalClass IITerminated10 Feb 2016
Z-0707-2016Allura Xper FD20; Model Numbers: 722028 722012 722006 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. "…13297 in totalClass IITerminated10 Feb 2016
Z-0708-2016UNIQ FD; Model Numbers: 722028 The Philips UNIQ is intended for use on human patients to perform: " Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as…13297 in totalClass IITerminated10 Feb 2016
Z-0709-2016UNIQ FD OR table; Model Number 722035 The Philips UNIQ is intended for use on human patients to perform: " Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal…13297 in totalClass IITerminated10 Feb 2016
Z-0710-2016Allura Xper FD20 OR Table; Model Numbers: 722035 722023 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. "…13297 in totalClass IITerminated10 Feb 2016
Z-0711-2016Allura Xper FD20 Biplane; Model Numbers: 722013 722008 722015 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis.…13297 in totalClass IITerminated10 Feb 2016
Z-0712-2016Allura Xper FD20 Biplane OR Table; Model Numbers: 722025 722020 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and…13297 in totalClass IITerminated10 Feb 2016
Z-0713-2016Allura Xper FD20/20; Model Number: 722038 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. " Cardiac imaging…13297 in totalClass IITerminated10 Feb 2016
Z-0714-2016Allura Xper FD20/20 OR Table; Model Number: 722039 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. " Cardiac…13297 in totalClass IITerminated10 Feb 2016
Z-0715-2016Allura CV20; Model Number: 722031 The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable internal diseases or…13297 in totalClass IITerminated10 Feb 2016
Z-0716-2016Cardiovascular Allura Centron; Model Number: 722400 The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable…13297 in totalClass IITerminated10 Feb 2016
Z-0717-2016INTEGRIS H3000; Model Number: 72238 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0718-2016INTEGRIS HM3000; Model Number: 72239 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0719-2016INTEGRIS BN/BV3000; Model Number: 72240 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS…13297 in totalClass IITerminated10 Feb 2016
Z-0720-2016INTEGRIS BH3000; Model Number: 72242 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0721-2016INTEGRIS V3000; Model Number: 72243, 72244, 72245 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips…13297 in totalClass IITerminated10 Feb 2016
Z-0722-2016INTEGRIS H5000; Model Number: 72246 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0723-2016INTEGRIS BH5000; Model Number: 72247 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0724-2016INTEGRIS V5000: Model Number: 72248 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0725-2016INTEGRIS BV5000: Model Number: 72249 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura…13297 in totalClass IITerminated10 Feb 2016
Z-0726-2016INTEGRIS Allura 9 (biplane); Model Number: 722021 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips…13297 in totalClass IITerminated10 Feb 2016
Z-0727-2016INTEGRIS Allura 15-12 (mono); Model Number: 722043 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips…13297 in totalClass IITerminated10 Feb 2016
Z-0728-2016INTEGRIS Allura 15-12 (biplane); Product Code: 722044 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips…13297 in totalClass IITerminated10 Feb 2016
Z-0729-2016INTEGRIS CV; Model Number: 722030 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura 12…13297 in totalClass IITerminated10 Feb 2016
Z-0730-2016INTEGRIS BV3000 MONO; Model Number: 72241 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS…13297 in totalClass IITerminated10 Feb 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
29Median of the firm: 1
Events of the firm on the register
61

Other events of the firm

Other FDA enforcement events of Philips Electronics North America Corporation
InitiatedProduct (first record)TypeClassRecords
15 Nov 2018Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 4535DeviceClass II1
10 Oct 2018IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health DeviceClass II1
16 Jul 2018INTEGRIS Allura 9, System Code 722018 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging aDeviceClass II27
13 Jul 2018Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neoDeviceClass II5
3 May 2018HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillatoDeviceClass II1
30 Apr 2018IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865350DeviceClass II4
16 Mar 2018Multiva systems, 1) Multiva 1.5T 8 R5 Model 781072, 2) Multiva 1.5T 16 R5 Model 781073, 3) Multiva 1.5T 8ch R5 based ModDeviceClass II13
14 Mar 2018Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray DeviceClass II1
7 Feb 2018HeartStart XL+ Defibrillator/Monitor The HeartStart XL+ is intended for use in a hospital setting by qualified medical pDeviceClass III1
29 Jan 2018Philips QCPR Meter - used with FR3 Product Usage: The Philips CPR meter is intended for use with the Philips HeartStart DeviceClass II2

Philips Electronics North America Corporation: all 61 events

Official channels

Cite this page

Recall Register. FDA enforcement event 72599. https://recallregister.com/fda/ii/72599/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and…
What quantity of product is in FDA event 72599?
The event has 29 product records. The first record gives this quantity: 13297 in total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
91% of the 1,153 device events that started in 2015 are in Class II.