Device recall: MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for…

Class II FDA enforcement event 72352. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic MiniMed Inc. started this device recall on 25 Sep 2015. The firm address in the FDA record is in Northridge, CA. FDA put the recall in Class II. FDA event 72352 has 2 product records. The first FDA enforcement report date is 23 Dec 2015. The record gives the status Terminated, with the termination date 5 Feb 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya,…

Recalling firm

Medtronic MiniMed Inc., Northridge, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 72352: product records
Recall numberProductQuantityClassStatusReport date
Z-0437-2016MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous…53 unitsClass IITerminated23 Dec 2015
Z-0438-2016MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is…1052 unitsClass IITerminated23 Dec 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
23

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 72352. https://recallregister.com/fda/ii/72352/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya,…
What quantity of product is in FDA event 72352?
The event has 2 product records. The first record gives this quantity: 53 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
91% of the 1,153 device events that started in 2015 are in Class II.