Recall Register

Device recall: Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000) (proprietary name: AFF (Mark I)…

Class II FDA enforcement event 72331. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The device might fail to detect air in line while infusing the medication Venofer. This may occur when a drop of Venofer gets stuck in the tubing adjacent to the bubble detector and the Total Volume Limit is programmed higher than the actual bag volume.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

WalkMed Infusion, LLC started this device recall on 1 Oct 2015. The firm address in the FDA record is in Englewood, CO. FDA put the recall in Class II. FDA event 72331 has 1 product record. The first FDA enforcement report date is 18 Nov 2015. The record gives the status Terminated, with the termination date 12 Jul 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distributed in the states of AL, CA, CO, DE, FL, GA, IL, MA, MD, MI, MN, MS, NC, NJ, NM, NV, NY, OH, PA, TN, TX, and WA.

Recalling firm

WalkMed Infusion, LLC, Englewood, CO.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 72331: product records
Recall numberProductQuantityClassStatusReport date
Z-0242-2016Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000) (proprietary name: AFF (Mark I) Volumetric Infusion Pump). Used to pump fluids into a patient in a controlled manner.4695 unitsClass IITerminated18 Nov 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 2
Events of the firm on the register
4

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 72331. https://recallregister.com/fda/ii/72331/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distributed in the states of AL, CA, CO, DE, FL, GA, IL, MA, MD, MI, MN, MS, NC, NJ, NM, NV, NY, OH, PA, TN, TX, and WA.
What quantity of product is in FDA event 72331?
The FDA record gives this quantity: 4695 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?