Recall Register

Device recall: GE Healthcare, Discovery MR750w 3.0T. Product Usage: The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w…

Class II FDA enforcement event 72301. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

GE Healthcare started this device recall on 31 Aug 2015. The firm address in the FDA record is in Waukesha, WI. FDA put the recall in Class II. FDA event 72301 has 7 product records. The first FDA enforcement report date is 18 Nov 2015. The record gives the status Terminated, with the termination date 7 Sep 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA,…

Recalling firm

GE Healthcare, Waukesha, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

7 product records in the event.

FDA event 72301: product records
Recall numberProductQuantityClassStatusReport date
Z-0254-2016GE Healthcare, Discovery MR750w 3.0T. Product Usage: The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T Systems are whole body magnetic resonance scanners for diagnostic imaging.756Class IITerminated18 Nov 2015
Z-0255-2016GE Healthcare, SIGNA 3.0T. Product Usage: Magnetic resonance system for diagnostic imaging. The indications for use for the 3.0T Signa VH/i (Signa 3.0T MR System) Transmit/Receive Body Imaging Coil expands the imaging capability of the 3.0T Signa VH/i MR Imaging System. The Transmit/Receive Body Imaging Coil is…19Class IITerminated18 Nov 2015
Z-0256-2016GE Healthcare, SIGNA Excite 3.0T. MR System for use as a diagnostic imaging device.55Class IITerminated18 Nov 2015
Z-0257-2016GE Healthcare, SIGNA HD 3.0T. Product Usage: The Signa Profile EXCITE MR system is an open, whole body scanner for use as a diagnostic imaging device. It may be utilized for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR safe biopsy needles.97Class IITerminated18 Nov 2015
Z-0258-2016GE Healthcare, SIGNA HDx 3.0T. Product Usage: The GE Signa HDx MR system is a whole body magnetic resonance scanner for use as a diagnostic imaging device.350Class IITerminated18 Nov 2015
Z-0259-2016GE Healthcare, SIGNA HDxt 3.0T. Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner for use as a diagnostic imaging device.215Class IITerminated18 Nov 2015
Z-0260-2016GE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high-resolution physiologic and anatomic information, acquired simultaneously and isocentrically.17Class IITerminated18 Nov 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class II
91%
Product records in this event
7Median of the firm: 1
Events of the firm on the register
44

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 72301. https://recallregister.com/fda/ii/72301/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA,…
What quantity of product is in FDA event 72301?
The event has 7 product records. The first record gives this quantity: 756
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
91% of the 1,153 device events that started in 2015 are in Class II.