Recall Register

Food, drugs, devices › Hardy Diagnostics

Device recall: HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/s

FDA enforcement event 71804 · Class II · initiated 9 Jul 2015 · status Terminated · terminated 10 Nov 2015

DeviceVoluntary: Firm initiated1 product recordSanta Maria, CA

FDA enforcement event 71804 is a device recall by Hardy Diagnostics of Santa Maria, CA, initiated 9 Jul 2015 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 10 Nov 2015. Initiation: voluntary: firm initiated. Hardy Diagnostics has 14 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Hardy Diagnostics, Santa Maria, CA.

Reason for recall

"Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot."

Distribution

"Distributed in the states of VA, PA, NY, NJ, IL, OH, and MD."

Product records

Recall numberProductQuantityClassStatusReport date
Z-2463-2015HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light Sensitive. In vitro diagnostic www.HardyDiagnostics.com2,820 (282 pk/10)Class IITerminated2 Sep 2015

Context

91%of 2015 device events are Class II1,153 events that year
1product record in this eventfirm median: 1
14events on this register for the firm

Other enforcement events for Hardy Diagnostics

InitiatedProduct (first record)TypeClassRecords
8 Jun 2026HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion suscDeviceClass II1
6 May 2025HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. DeviceClass II1
16 Apr 2021Quickslide GramPro 1 Automated Gram Stanier is a microscope slide stainer.DeviceClass II1
21 Dec 2020Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured: 2020-10-29 Expires: 2020-12-2DeviceClass II1
23 Oct 2020Viral Transport Medium, 3ml, Ref: R99, Container type: Tube, Packaged: 20 tubes/box, IVD, CVM Transport Media is provideDeviceClass II1
6 Oct 2020Strep B Carrot Broth One-Step, Ref. Z40, UDI: 00819398026499DeviceClass III1
14 Nov 2019HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium for the qualitative direct detecDeviceClass II1
24 May 2016QuickSlide MGS-80, Automated Mutli-Slide Gram Stainer/GramPRO-80. Systematically stains, rinses, decolorizes, and counteDeviceClass II1
21 Mar 2016Hardy Diagnostics GBS Detect Cat no. A300. Used for the isolation and detection by enhanced hemolysis of gamma-hemolyticDeviceClass II1
18 Nov 2015FB Broth, 10ml (Fastidious Bacteria Broth) Cat. no: K31 Lot: 15231 A fastidious bacteria broth for the enriched cultivatDeviceClass II1

All 14 events for Hardy Diagnostics

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Distributed in the states of VA, PA, NY, NJ, IL, OH, and MD."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 10 Nov 2015. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.