Recall Register

Food, drugs, devices › Maquet Cardiovascular Us Sales, Llc

Device recall: Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH &

FDA enforcement event 71446 · Class II · initiated 10 Jun 2015 · status Terminated · terminated 18 Apr 2016

DeviceVoluntary: Firm initiated1 product recordWayne, NJ

FDA enforcement event 71446 is a device recall by Maquet Cardiovascular Us Sales, Llc of Wayne, NJ, initiated 10 Jun 2015 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 18 Apr 2016. Initiation: voluntary: firm initiated. Maquet Cardiovascular Us Sales, Llc has 40 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Maquet Cardiovascular Us Sales, Llc, Wayne, NJ.

Reason for recall

"There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision."

Distribution

"Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-2097-2015Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.12 unitsClass IITerminated29 Jul 2015

Context

91%of 2015 device events are Class II1,153 events that year
1product record in this eventfirm median: 1
40events on this register for the firm

Other enforcement events for Maquet Cardiovascular Us Sales, Llc

InitiatedProduct (first record)TypeClassRecords
20 Nov 2020HLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 - Product UsagDeviceClass II2
30 Sep 2020Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonaDeviceClass II1
24 Sep 2020Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during anDeviceClass II2
7 Aug 2020Maquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - Centrifugal Pump SysteDeviceClass III1
18 Jun 2020Holder HKHZ 19 for use with Quadrox-i and Quadrox-iD Neonatal / Pediatric Oxygenators, Model Number: 701047495 - ProductDeviceClass II1
20 Feb 2020Avalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039DeviceClass II1
10 Feb 2020Getinge 88-Series-washer disinfectors intended for cleaning, disinfection and drying Model Number: 88-5 Part Number: 88-DeviceClass II1
16 Jan 2020Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as CentraDeviceClass III2
3 Dec 2019Maquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW SDeviceClass II1
11 Jul 2019Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794DeviceClass II1

All 40 events for Maquet Cardiovascular Us Sales, Llc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 18 Apr 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.