Device recall: MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting…
Class II FDA enforcement event 70893. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Beckman Coulter, Inc. started this device recall on 8 Aug 2014. The firm address in the FDA record is in West Sacramento, CA. FDA put the recall in Class II. FDA event 70893 has 7 product records. The first FDA enforcement report date is 6 May 2015. The record gives the status Terminated, with the termination date 27 Apr 2015.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian…
Recalling firm
Beckman Coulter, Inc., West Sacramento, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
7 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1528-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg Combo 50, Catalog B1017-406 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1529-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B1017-417 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels.… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1530-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 51, Catalog B1017-407 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels.… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1531-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 55, Catalog B1017-409 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels.… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1532-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 61, Catalog B1017-414 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels.… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1533-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg MIC 38, Catalog B1017-412 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
| Z-1534-2015 | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: ES L plus, Catalog B1027-101 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan… | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total | Class II | Terminated | 6 May 2015 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2014
- 1,176
- Share of these events in Class II
- 91%
- Product records in this event
- 7Median of the firm: 1
- Events of the firm on the register
- 216
Other events of the firm
Official channels
- FDA enforcement report Z-1528-2015The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 70893. https://recallregister.com/fda/ii/70893/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian…
- What quantity of product is in FDA event 70893?
- The event has 7 product records. The first record gives this quantity:
Approx ~(216240 boxes U.S. and 84249 boxes OUS) total
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2014 have the same class?
- 91% of the 1,176 device events that started in 2014 are in Class II.