Device recall: Brand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a…
Class II FDA enforcement event 70676. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Zimmer, Inc is voluntarily recalling 22 lots of Persona Anterior Referencing Sizer with Locking Boom, part number 42-5099-088-10. Devices are being removed from distribution due to nonconforming to print specifications.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Zimmer, Inc. started this device recall on 10 Mar 2015. The firm address in the FDA record is in Warsaw, IN. FDA put the recall in Class II. FDA event 70676 has 1 product record. The first FDA enforcement report date is 8 Apr 2015. The record gives the status Terminated, with the termination date 6 Nov 2015.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - US including Domestic: Puerto Rico, AZ, NY. OR, ME, NC, OH, TX, CO, IL, TN, PA, HI, CA, KS, WI, VA, MN, NJ, UT, MA, GA, MI, OK, MS, FL, AL, IN, and Internationally to…
Recalling firm
Zimmer, Inc., Warsaw, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1360-2015 | Brand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. The Persona Surgical Technique (97-5026-001-00, Rev. 9) states that the sizer helps ensure that the desired external… | 688 | Class II | Terminated | 8 Apr 2015 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2015
- 1,153
- Share of these events in Class II
- 91%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 108
Other events of the firm
Official channels
- FDA enforcement report Z-1360-2015The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 70676. https://recallregister.com/fda/ii/70676/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - US including Domestic: Puerto Rico, AZ, NY. OR, ME, NC, OH, TX, CO, IL, TN, PA, HI, CA, KS, WI, VA, MN, NJ, UT, MA, GA, MI, OK, MS, FL, AL, IN, and Internationally to…
- What quantity of product is in FDA event 70676?
- The FDA record gives this quantity:
688
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2015 have the same class?
- 91% of the 1,153 device events that started in 2015 are in Class II.