Recall Register

Device recall: Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A

Class II FDA enforcement event 69712. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Philips Medical Systems, Inc. started this device recall on 19 Nov 2014. The firm address in the FDA record is in Andover, MA. FDA put the recall in Class II. FDA event 69712 has 1 product record. The first FDA enforcement report date is 11 Nov 2015. The record gives the status Terminated, with the termination date 15 Aug 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy,…

Recalling firm

Philips Medical Systems, Inc., Andover, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 69712: product records
Recall numberProductQuantityClassStatusReport date
Z-0204-2016Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A1553 unitsClass IITerminated11 Nov 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
47

Other events of the firm

Other FDA enforcement events of Philips Medical Systems, Inc.
InitiatedProduct (first record)TypeClassRecords
23 Apr 2015Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medicalDeviceClass II1
9 Feb 2015Philips Healthcare DigitalDiagnost System X-RayDeviceClass II2
3 Feb 2015Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21 Allura XpeDeviceClass II9
23 Jan 2015Philips DuraDiagnost X-Ray system, Stationary X-Ray System Radiographic Models: 712210, 712211DeviceClass II1
23 Dec 2014HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 DeviceClass II1
19 Dec 2014Philips IntelliVue Information Center iX (release A.00, A.01, and A.02) are impacted by this issue: 866023 IntelliVue InDeviceClass II1
20 Nov 2014Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065 and 9067. DeviceClass II1
19 Nov 2014Philips HeartStart MRx Monitor/Defibrillator Models:M3535A, M3536A, M3536M, M3536MC, M3536M4, M3536M5 The HeartStart MRxDeviceClass II1
19 Nov 2014Philips IntelliVue TcG10; Monitor, carbon-dioxide, cutaneous. The IntelliVue TcG10 is a device for the measurement of thDeviceClass II1
19 Nov 2014HeartStart MRx Monitor/Defibrillator; Model. Numbers: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5DeviceClass II1

Philips Medical Systems, Inc.: all 47 events

Official channels

Cite this page

Recall Register. FDA enforcement event 69712. https://recallregister.com/fda/ii/69712/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy,…
What quantity of product is in FDA event 69712?
The FDA record gives this quantity: 1553 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?