Recall Register

Device recall: Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitro diagnostic…

Class II FDA enforcement event 69191. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Dako North America Inc. started this device recall on 29 Aug 2014. The firm address in the FDA record is in Carpinteria, CA. FDA put the recall in Class II. FDA event 69191 has 1 product record. The first FDA enforcement report date is 1 Oct 2014. The record gives the status Terminated, with the termination date 3 Mar 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US Nationwide in the states of AT, AU, BE, CA, CN, DE, DK, ES, Fl, FR, GB, IE, IT, NL, NO, SE, US, DZ, GR, IS, OM, SG, UY.

Recalling firm

Dako North America Inc., Carpinteria, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 69191: product records
Recall numberProductQuantityClassStatusReport date
Z-2708-2014Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitro diagnostic use. The Dako Autostainer Link 48 is an automated slide processing system compatible with reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine…191 units total (57 units in US)Class IITerminated1 Oct 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
8

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 69191. https://recallregister.com/fda/ii/69191/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US Nationwide in the states of AT, AU, BE, CA, CN, DE, DK, ES, Fl, FR, GB, IE, IT, NL, NO, SE, US, DZ, GR, IS, OM, SG, UY.
What quantity of product is in FDA event 69191?
The FDA record gives this quantity: 191 units total (57 units in US)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.