Device recall: ADVIA Centaur Immunoassay System (including refurbished units), This system is intended for professional use in a…

Class II FDA enforcement event 69084. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm has identified an issue with patient demographic information sent to the LIS from the ADVIA Centaur/ADVIA Centaur XP Immunoassay systems. Siemens has confirmed that under extremely rare circumstances patient demographic data from the previous order received from the LIS is merged with the next order. This issue can occur when the LIS data buffer on the ADVIA Centaur system becomes full

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Healthcare Diagnostics started this device recall on 19 Aug 2014. The firm address in the FDA record is in Tarrytown, NY. FDA put the recall in Class II. FDA event 69084 has 2 product records. The first FDA enforcement report date is 8 Oct 2014. The record gives the status Terminated, with the termination date 22 May 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution including the United States., and the countries of Argentina, Austria, Australia, Bahrain, Bangladesh, Belgium, Belgrade, Brazil, Botswana, Canada, Chile, China, Colombia,…

Recalling firm

Siemens Healthcare Diagnostics, Tarrytown, NY.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 69084: product records
Recall numberProductQuantityClassStatusReport date
Z-2715-2014ADVIA Centaur Immunoassay System (including refurbished units), This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. As with all diagnostic tests, a definitive clinical diagnosis should not be based on the results of…Domestic: 2771 units; Foreign: 7744 unitsClass IITerminated8 Oct 2014
Z-2716-2014ADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. As with all diagnostic tests, a definitive clinical diagnosis should not be based on the results…Domestic: 2771 units; Foreign: 7744 unitsClass IITerminated8 Oct 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
17

Other events of the firm

Other FDA enforcement events of Siemens Healthcare Diagnostics
InitiatedProduct (first record)TypeClassRecords
17 Oct 2017BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184DeviceClass II1
19 Sep 2014IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use wDeviceClass II1
13 Aug 2014ADVIA Centaur Immunoassay System, Catalog Numbers/Siemens Material Numbers (SMN) 10284980, 10286140, 10309524, 10309525,DeviceClass II2
6 Aug 2014ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material NumbeDeviceClass II1
18 Feb 2014ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XPDeviceClass II1
20 Dec 2013IMMULITE /IMMULITE 1000 Systems TIE Total IgE, REF LKIE1, SMN 10380867, IVD --- Siemens Healthcare Diagnostics Products DeviceClass II2
20 Dec 2013IMMULITE/IMMULITE 1000 Systems Androstenedione (Test Code AND), REF LKAO1, SMN 10381158, IVD, 100 test kit size --- SiemDeviceClass II2
6 Nov 2013IMMULITE/IMMULITE 1000 Systems FOL Folic Acid, REF/Catalog Number LKFO1 (100 tests), Siemens Material Number (SMN) 10380DeviceClass II1
24 Sep 2013Siemens Chemistry Calibrator (CAL) used with the ADVIA Chemistry Direct Bilirubin_2 Reagent, Siemens Material Number (SMDeviceClass II1
3 Sep 2013ADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and specialty chemistries, methods for speciDeviceClass II4

Siemens Healthcare Diagnostics: all 17 events

Official channels

Cite this page

Recall Register. FDA enforcement event 69084. https://recallregister.com/fda/ii/69084/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution including the United States., and the countries of Argentina, Austria, Australia, Bahrain, Bangladesh, Belgium, Belgrade, Brazil, Botswana, Canada, Chile, China, Colombia,…
What quantity of product is in FDA event 69084?
The event has 2 product records. The first record gives this quantity: Domestic: 2771 units; Foreign: 7744 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.