Device recall: The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of t

FDA enforcement event 69045 · Class II · initiated 1 Sep 2011 · status Terminated · terminated 1 Jul 2015

DeviceVoluntary: Firm initiated1 product recordCleveland, OH

FDA enforcement event 69045 is a device recall by Philips Medical Systems (Cleveland) Inc of Cleveland, OH, initiated 1 Sep 2011 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 89% of the 114 device enforcement events initiated in 2011 carry the same class.

Status: Terminated, terminated 1 Jul 2015. Initiation: voluntary: firm initiated. Philips Medical Systems (Cleveland) Inc has 129 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Philips Medical Systems (Cleveland) Inc, Cleveland, OH.

Reason for recall

"During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software. When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images. Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software. Philips will issue a software up"

Distribution

"Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden."

Product records

Recall numberProductQuantityClassStatusReport date
Z-2664-2014The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports,…6Class IITerminated24 Sep 2014

Context

89%of 2011 device events are Class II114 events that year
1product record in this eventfirm median: 1
129events on this register for the firm

Other enforcement events for Philips Medical Systems (Cleveland) Inc

InitiatedProduct (first record)TypeClassRecords
5 Aug 2025Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)0088483DeviceClass II1
17 Jul 2025Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System SoftwareDeviceClass II1
30 Aug 2022Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.DeviceClass II1
19 Sep 2019Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positDeviceClass I5
24 Jun 2019Ingenuity CT, Model # 728326, computed tomography x-ray systemDeviceClass II9
10 Jun 2019BrightView SPECT, Model Number 882480DeviceClass II3
3 Jun 2019Vereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) sysDeviceClass II1
19 Dec 2018Ingenuity TF PET/CT (non-RoHS), Model Number 882456 This device is a diagnostic imaging system that combines Positron EmDeviceClass II1
18 Oct 2018Brilliance 64, Model 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images DeviceClass II6
4 Sep 2018Vereos PET/CT, Software version 2.0.1. Model Number: 882446 Product Usage: The Philips Vereos PET/CT system is a diagnosDeviceClass II1

All 129 events for Philips Medical Systems (Cleveland) Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software. When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images. Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software. Philips will issue a software up"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 1 Jul 2015. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.