Device recall: Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation the
FDA enforcement event 68985 is a device recall by Elekta, Inc. of Atlanta, GA, initiated 18 Aug 2014 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.
Status: Terminated, terminated 15 Sep 2021. Initiation: voluntary: firm initiated. Elekta, Inc. has 105 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Elekta, Inc., Atlanta, GA.
Reason for recall
"Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan."
Distribution
"Worldwide distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, FL, IL, IN, MD, MA, MI, MN, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, Puerto Rico;…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2581-2014 | Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given… | 928 units | Class II | Terminated | 10 Sep 2014 |
Context
Other enforcement events for Elekta, Inc.
All 105 events for Elekta, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, FL, IL, IN, MD, MA, MI, MN, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, Puerto Rico;…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 15 Sep 2021. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.