Food, drugs, devices › Hospira Inc.
Device recall: The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS
Hospira Inc. (Lake Forest, IL) initiated this device recall on 6 Jun 2014; FDA lists 1 product record under event 68658, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.
Status: Terminated, terminated 6 Dec 2016. Initiation: voluntary: firm initiated. Hospira Inc. has 151 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Hospira Inc., Lake Forest, IL.
Reason for recall
"Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a"
Distribution
"US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2027-2014 | The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send,… | 13 units | Class II | Terminated | 16 Jul 2014 |
Context
Other enforcement events for Hospira Inc.
All 151 events for Hospira Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 6 Dec 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.