Recall Register

Food, drugs, devices › Zimmer, Inc.

Device recall: PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilizat

FDA enforcement event 68653 · Class II · initiated 1 Jul 2014 · status Terminated · terminated 9 Jun 2015

DeviceVoluntary: Firm initiated2 product recordsWarsaw, IN

FDA enforcement event 68653 is a device recall by Zimmer, Inc. of Warsaw, IN, initiated 1 Jul 2014 and classified Class II, with 2 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.

Status: Terminated, terminated 9 Jun 2015. Initiation: voluntary: firm initiated. Zimmer, Inc. has 108 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Zimmer, Inc., Warsaw, IN.

Reason for recall

"The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide."

Distribution

"Worldwide Distribution - USA including AZ, CA, CO, FL, IL, IN, KS, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX & VA and Internationally to AUSTRALIA, FRANCE, INDIA, ITALY, and GERMANY."

Product records

Recall numberProductQuantityClassStatusReport date
Z-2288-2014PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during…190 unitsClass IITerminated27 Aug 2014
Z-2289-2014PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during…190 unitsClass IITerminated27 Aug 2014

Context

91%of 2014 device events are Class II1,176 events that year
2product records in this eventfirm median: 1
108events on this register for the firm

Other enforcement events for Zimmer, Inc.

InitiatedProduct (first record)TypeClassRecords
20 May 2026Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240DeviceClass II5
11 May 2026Brand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/Catalog NumbDeviceClass II1
30 Mar 2026Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011DeviceClass II1
2 Dec 2025Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number:DeviceClass II17
26 Sep 2025Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/CaDeviceClass II2
24 Mar 2025Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-ConstrDeviceClass II5
27 Feb 2025NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model NumberDeviceClass II1
12 Feb 2025Z1 Femoral Hip System, Product Number 611777612DeviceClass II1
6 Jan 2025NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14DeviceClass II2
17 Sep 2024Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access foDeviceClass II1

All 108 events for Zimmer, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - USA including AZ, CA, CO, FL, IL, IN, KS, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX & VA and Internationally to AUSTRALIA, FRANCE, INDIA, ITALY, and GERMANY."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 9 Jun 2015. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.