Device recall: The LifeCare PCA Plus II infusion pump system is designed for safe, effective, analgesic delivery to patients through…

Class II FDA enforcement event 68318. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Hospira Inc. started this device recall on 8 Mar 2013. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class II. FDA event 68318 has 4 product records. The first FDA enforcement report date is 20 Jul 2016. The record gives the status Terminated, with the termination date 7 Nov 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,…

Recalling firm

Hospira Inc., Lake Forest, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 68318: product records
Recall numberProductQuantityClassStatusReport date
Z-2139-2016The LifeCare PCA Plus II infusion pump system is designed for safe, effective, analgesic delivery to patients through the intravenous or epidural routes in a wide range of clinical settings, including ICU, CCU, SICU, trauma units, bone marrow transplant centers, burn units, oncology centers, and medical/surgical…53,151 infusion pump systemsClass IITerminated20 Jul 2016
Z-2140-2016The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to…8,216 infusion pump systemsClass IITerminated20 Jul 2016
Z-2141-2016LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of clinical settings that…38,878 infusion pump systemsClass IITerminated20 Jul 2016
Z-2142-2016The LifeCare Pump Model 4200 PCA Infuser makes possible the practical implementation of patient-controlled analgesia. The health care professional sets the prescribed dose, rates, and limits for the drug to be infused. The infuser can be set to operate in one of three modes: PCA Mode, Continuous Mode, and PCA +…6,268 infusion pump systemsClass IITerminated20 Jul 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
151

Other events of the firm

Other FDA enforcement events of Hospira Inc.
InitiatedProduct (first record)TypeClassRecords
21 Apr 2017Infant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, Rx only, HospirDrugClass I1
1 Nov 2016Fentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., Lake Forest IL 6DrugClass II1
15 Aug 2016DOBUTamine Injection, USP 250 mg/20 mL, 20 mL Vial, For IV Use Only, Must be Diluted Prior to Use, (12.5 mg per mL), Rx DrugClass II1
29 Jun 20165% Lidocaine HCL and 7.5% Dextrose Injection, USP, 2 mL Single-dose ampule, 5 count box, Rx Only, For Spinal Anesthesia DrugClass III1
23 Jun 2016Diazepam Injection USP 10mg/2mL (5 mg/mL, 2 mL), Rx only, Hospira Inc. Lake Forest, IL 60045, NDC 0409-1273-32DrugClass II1
6 May 2016Vancomycin Hydrochloride for Injection, USP, 10 g, 100 mL vial, Rx only, Hospira, Lake Forest, IL, NDC 0409-6510-01DrugClass II1
25 Mar 2016The Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enterDeviceClass II2
23 Mar 2016Magnesium Sulfate Inj., USP 50%, 10 g/20 mL (0.5 g/mL), (4 mEq Magnesium/mL, 20 mL Single-dose vial, packaged in 25 vialDrugClass III1
23 Mar 2016Magnesium Sulfate Inj., USP 50%, 10 g/20 mL (0.5 g/mL), (4 mEq Magnesium/mL, 20 mL Single-dose vial, packaged in 25 vialDrugClass I1
18 Mar 20168.4% Sodium Bicarbonate Injection, USP 50mEq (1mEq/mL), 4.2 grams (84 mg/mL), 50 mL, Rx only, Hospira Inc, Lake Forest, DrugClass I1

Hospira Inc.: all 151 events

Official channels

Cite this page

Recall Register. FDA enforcement event 68318. https://recallregister.com/fda/ii/68318/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,…
What quantity of product is in FDA event 68318?
The event has 4 product records. The first record gives this quantity: 53,151 infusion pump systems
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.