Device recall: Medtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary Pacemaker…

Class II FDA enforcement event 68165. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Events related to a pacing rate outside of the intended setting, including events of sudden increased pacing rate up to the maximum setting of 180 pulses per minute (ppm), have been reported. This is caused by the development of high resistance on internal electrical connector contacts due to oxidation over time.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Inc. Cardiac Rhythm Disease Management started this device recall on 29 Apr 2014. The firm address in the FDA record is in Saint Paul, MN. FDA put the recall in Class II. FDA event 68165 has 1 product record. The first FDA enforcement report date is 4 Jun 2014. The record gives the status Terminated, with the termination date 8 Oct 2014.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution -- USA, including the territories of Guam and Puerto Rico; and, the countries of Albania, Algeria, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Benin, Bolivia,…

Recalling firm

Medtronic Inc. Cardiac Rhythm Disease Management, Saint Paul, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 68165: product records
Recall numberProductQuantityClassStatusReport date
Z-1661-2014Medtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary Pacemaker (AAI/VVI). Used in conjunction with a cardiac pacing lead system for temporary atrial or ventricular pacing in a clinical environment. The device is typically connected to temporary transvenous, epicardial or…16,337 (8,989 US; 7,348 OUS)Class IITerminated4 Jun 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
16

Other events of the firm

Other FDA enforcement events of Medtronic Inc. Cardiac Rhythm Disease Management
InitiatedProduct (first record)TypeClassRecords
15 Sep 2015Medtronic Medical Device Identification Card for SureScan pacemaker patients. The front of this card identifies the implDeviceClass II1
1 Sep 2015Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.DeviceClass II1
23 Mar 2015Covidien Trellis 6 and Trellis 8 Peripheral Infusion Systems, Sterile EO, Rx only. Models BVT608010V01, BVT608030V01, BVDeviceClass II1
17 Mar 2015Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cDeviceClass II1
23 Feb 2015Medtronic, 10 Fr FlexCath Select Steerable Sheath 990065. The FlexCath Select Steerable Sheath (10 Fr Select) is used foDeviceClass I1
30 Jan 2014Medtronic Select 3D Arterial Cannulae. Vented, models 78618 (18 Fr.) and 78620 (20 Fr.). Non-Vented Cannulae, models 787DeviceClass II6
17 Jul 2013Medtronic Custom Perfusion System. Model Numbers: 1D80R8, 1E34R1, 1F75R1, 1G49R2, 1G75R1, 1P07R8, 1P91R4, 5B15R11, and 5DeviceClass II3
8 Jul 2013Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters into the lDeviceClass II1
14 Jun 2013Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/6/13:DeviceClass II1
12 Jun 2013Medtronic, Launcher Guiding Catheter, Model # LA8EBU45SH, Lot # 0006722866, 8F and Model # LA6IMAD, Lot # 0006724224, 6FDeviceClass II1

Medtronic Inc. Cardiac Rhythm Disease Management: all 16 events

Official channels

Cite this page

Recall Register. FDA enforcement event 68165. https://recallregister.com/fda/ii/68165/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution -- USA, including the territories of Guam and Puerto Rico; and, the countries of Albania, Algeria, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Benin, Bolivia,…
What quantity of product is in FDA event 68165?
The FDA record gives this quantity: 16,337 (8,989 US; 7,348 OUS)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.