Reason for recall, in the words of FDA

Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Roche Diagnostics Operations, Inc. started this device recall on 28 Feb 2014. The firm address in the FDA record is in Indianapolis, IN. FDA put the recall in Class II. FDA event 67549 has 1 product record. The first FDA enforcement report date is 9 Apr 2014. The record gives the status Terminated, with the termination date 10 Jan 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA nationwide including PR and the country of Guam.

Recalling firm

Roche Diagnostics Operations, Inc., Indianapolis, IN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 67549: product records
Recall numberProductQuantityClassStatusReport date
Z-1315-2014Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of the following systems: 1. 03337103001, cobas b 221<1>Roche OMNI S1 system 2. 03337111001, cobas b 221<2>Roche OMNI S2 system 3. 03337154001,cobas b 221<6>Roche OMNI S6 system 4. 03337138001, cobas b…N/AClass IITerminated9 Apr 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
69

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 67549. https://recallregister.com/fda/ii/67549/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA nationwide including PR and the country of Guam.
What quantity of product is in FDA event 67549?
The FDA record gives this quantity: N/A
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.