Recall Register

Device recall: LifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by…

Class II FDA enforcement event 67453. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Hospira Inc. started this device recall on 24 Jan 2014. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class II. FDA event 67453 has 3 product records. The first FDA enforcement report date is 11 Jun 2014. The record gives the status Terminated, with the termination date 7 Nov 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada,…

Recalling firm

Hospira Inc., Lake Forest, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 67453: product records
Recall numberProductQuantityClassStatusReport date
Z-1702-2014LifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by intravenous (I.V.) administration.List Number 02507: 132,889 pumpsClass IITerminated11 Jun 2014
Z-1703-2014Plum XL, Micro Macro Plum XL3 with data port, and Micro Macro Plum XL with data port infusion pumps, indicated for intravenous, arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products to patients in hospital and home care…List Number 11555: 17,887 pumps; List Number 11855: 1,292 pumps; List Number 11859: 147,399 pumpsClass IITerminated11 Jun 2014
Z-1704-2014Plum XL3 Multi line Infusion System, Micro Macro Plum XL3 Multi line Infusion System, and Micro Macro Plum XL Infusion System. Infusion pumps for intravenous. arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products to patients in…List Number 11781: 1,828 pumps; List Number 11845: 13,700 pumps; List Number 11846: 22,671 pumpsClass IITerminated11 Jun 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class II
91%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
151

Other events of the firm

Other FDA enforcement events of Hospira Inc.
InitiatedProduct (first record)TypeClassRecords
21 Apr 2017Infant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, Rx only, HospirDrugClass I1
1 Nov 2016Fentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., Lake Forest IL 6DrugClass II1
15 Aug 2016DOBUTamine Injection, USP 250 mg/20 mL, 20 mL Vial, For IV Use Only, Must be Diluted Prior to Use, (12.5 mg per mL), Rx DrugClass II1
29 Jun 20165% Lidocaine HCL and 7.5% Dextrose Injection, USP, 2 mL Single-dose ampule, 5 count box, Rx Only, For Spinal Anesthesia DrugClass III1
23 Jun 2016Diazepam Injection USP 10mg/2mL (5 mg/mL, 2 mL), Rx only, Hospira Inc. Lake Forest, IL 60045, NDC 0409-1273-32DrugClass II1
6 May 2016Vancomycin Hydrochloride for Injection, USP, 10 g, 100 mL vial, Rx only, Hospira, Lake Forest, IL, NDC 0409-6510-01DrugClass II1
25 Mar 2016The Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enterDeviceClass II2
23 Mar 2016Magnesium Sulfate Inj., USP 50%, 10 g/20 mL (0.5 g/mL), (4 mEq Magnesium/mL, 20 mL Single-dose vial, packaged in 25 vialDrugClass III1
23 Mar 2016Magnesium Sulfate Inj., USP 50%, 10 g/20 mL (0.5 g/mL), (4 mEq Magnesium/mL, 20 mL Single-dose vial, packaged in 25 vialDrugClass I1
18 Mar 20168.4% Sodium Bicarbonate Injection, USP 50mEq (1mEq/mL), 4.2 grams (84 mg/mL), 50 mL, Rx only, Hospira Inc, Lake Forest, DrugClass I1

Hospira Inc.: all 151 events

Official channels

Cite this page

Recall Register. FDA enforcement event 67453. https://recallregister.com/fda/ii/67453/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada,…
What quantity of product is in FDA event 67453?
The event has 3 product records. The first record gives this quantity: List Number 02507: 132,889 pumps
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
91% of the 1,176 device events that started in 2014 are in Class II.