Device recall: Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, N
FDA enforcement event 67094 is a device recall by Integra LifeSciences Corp. of Plainsboro, NJ, initiated 26 Nov 2013 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 90% of the 1,236 device enforcement events initiated in 2013 carry the same class.
Status: Terminated, terminated 13 Feb 2015. Initiation: voluntary: firm initiated. Integra LifeSciences Corp. has 79 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ.
Reason for recall
"Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly."
Distribution
"Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0602-2014 | Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature… | 275 CAM01 Monitors | Class II | Terminated | 8 Jan 2014 |
Context
Other enforcement events for Integra LifeSciences Corp.
All 79 events for Integra LifeSciences Corp.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain,…"
Is the recall still ongoing?