Recall Register

Device recall: C-Series: Clinac, Trilogy, Trilogy Tx. , Novalis high energy linear accelerators and UNIQUE single energy linear…

Class II FDA enforcement event 66764. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

This correction is to notify users that a solution to a previous correction has been developed and Varians reps will be contacting locations to schedule installation.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Varian Medical Systems, Inc. started this device recall on 24 Oct 2013. The firm address in the FDA record is in Palo Alto, CA. FDA put the recall in Class II. FDA event 66764 has 1 product record. The first FDA enforcement report date is 11 Dec 2013. The record gives the status Terminated, with the termination date 13 Jan 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA Nationwide and in the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Belorussia, Bosnia and Herzegovina, Brazil, Bulgaria,…

Recalling firm

Varian Medical Systems, Inc., Palo Alto, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 66764: product records
Recall numberProductQuantityClassStatusReport date
Z-0439-2014C-Series: Clinac, Trilogy, Trilogy Tx. , Novalis high energy linear accelerators and UNIQUE single energy linear accelerator; Versions 7, 8 and 9. The UNIQUE is not sold in the US. Product Usage: The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for…3,809 devicesClass IITerminated11 Dec 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
19

Other events of the firm

Other FDA enforcement events of Varian Medical Systems, Inc.
InitiatedProduct (first record)TypeClassRecords
12 Aug 2019Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120 Multileaf CollDeviceClass II1
26 Dec 2018Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse TreatmentDeviceClass II1
2 Dec 2016Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.DeviceClass II1
13 Sep 2016Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head FrameDeviceClass II1
23 Jun 2016Optical Surface Monitoring System{OSMS, Varian Private Label} This is used on Varian's TrueBeam EDGE; Radiology: The AliDeviceClass II2
11 Jan 2016Eclipse Treatment Planning System versions 11, 13.0, 13.5 and 13.6; Model number: H48; Product Usage: The Eclipse TreatmDeviceClass II1
28 Sep 2015ARIA Radiation Oncology, versions 10, 11 , 13.0 and 13.5 with Clinical Assessment License. Model number HIT Product UsagDeviceClass II1
18 May 2015The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical DeviceClass II1
18 May 2015Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.DeviceClass II1
18 Aug 20144D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery bDeviceClass II1

Varian Medical Systems, Inc.: all 19 events

Official channels

Cite this page

Recall Register. FDA enforcement event 66764. https://recallregister.com/fda/ii/66764/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA Nationwide and in the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Belorussia, Bosnia and Herzegovina, Brazil, Bulgaria,…
What quantity of product is in FDA event 66764?
The FDA record gives this quantity: 3,809 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.