Device recall: FEM COND BUTTRESS PLT, RT.

Class II FDA enforcement event 66499. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Zimmer, Inc. started this device recall on 5 Nov 2013. The firm address in the FDA record is in Warsaw, IN. FDA put the recall in Class II. FDA event 66499 has 8 product records. The first FDA enforcement report date is 8 Jan 2014. The record gives the status Terminated, with the termination date 5 May 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution.

Recalling firm

Zimmer, Inc., Warsaw, IN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

8 product records in the event.

FDA event 66499: product records
Recall numberProductQuantityClassStatusReport date
Z-0585-2014FEM COND BUTTRESS PLT, RT.42, 065 units totalClass IITerminated8 Jan 2014
Z-0586-2014TIBIO FEM/FEM NAIL.42, 065 units totalClass IITerminated8 Jan 2014
Z-0587-2014DIST LAT FEM LOCK PLT, LT42, 065 units totalClass IITerminated8 Jan 2014
Z-0588-2014DIST LAT FEM LOCK PLT, RT42, 065 units totalClass IITerminated8 Jan 2014
Z-0589-2014ITST THREADED GUIDE PIN42, 065 units totalClass IITerminated8 Jan 2014
Z-0590-2014NGK Stem Ext 24 x 245mm42, 065 units totalClass IITerminated8 Jan 2014
Z-0591-2014FEM COND BUTTRESS PLT, LT42, 065 units totalClass IITerminated8 Jan 2014
Z-0592-2014TIB NAIL42, 065 units totalClass IITerminated8 Jan 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
8Median of the firm: 1
Events of the firm on the register
108

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 66499. https://recallregister.com/fda/ii/66499/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution.
What quantity of product is in FDA event 66499?
The event has 8 product records. The first record gives this quantity: 42, 065 units total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.