Recall Register

Device recall: Cordis OPTEASE Vena Cava Filter Is indicated for use in the prevention of recurrent pulmonary embolism (PE) via…

Class II FDA enforcement event 66483. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Cordis identified a printing error on one unit of the Optease Vena Cava Filter, in which the orientation arrow for the femoral approach was printed in the incorrect direction. The error resulted in the filter being implanted upside down, requiring and additional percutaneous procedure to retrieve the filter.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Cordis Corporation started this device recall on 27 Sep 2013. The firm address in the FDA record is in Miami Lakes, FL. FDA put the recall in Class II. FDA event 66483 has 1 product record. The first FDA enforcement report date is 6 Nov 2013. The record gives the status Terminated, with the termination date 23 Sep 2014.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US (nationwide) including Hawaii and Puerto Rico and states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND,…

Recalling firm

Cordis Corporation, Miami Lakes, FL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 66483: product records
Recall numberProductQuantityClassStatusReport date
Z-0081-2014Cordis OPTEASE Vena Cava Filter Is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in the following situations: - Pulmonary thromboembolism when anticoagulants are contraindicated, - Failure of anticoagulant therapy for thromboembolitic disease, -…29,395 unitsClass IITerminated6 Nov 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
23

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 66483. https://recallregister.com/fda/ii/66483/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US (nationwide) including Hawaii and Puerto Rico and states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND,…
What quantity of product is in FDA event 66483?
The FDA record gives this quantity: 29,395 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.