Recall Register

Food, drugs, devices › Intuitive Surgical, Inc.

Device recall: 8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosur

FDA enforcement event 65194 · Class II · initiated 8 May 2013 · status Terminated · terminated 18 Feb 2014

DeviceVoluntary: Firm initiated3 product recordsSunnyvale, CA

Intuitive Surgical, Inc. (Sunnyvale, CA) initiated this device recall on 8 May 2013; FDA lists 3 product records under event 65194, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 90% of the 1,236 device enforcement events initiated in 2013 carry the same class.

Status: Terminated, terminated 18 Feb 2014. Initiation: voluntary: firm initiated. Intuitive Surgical, Inc. has 105 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Intuitive Surgical, Inc., Sunnyvale, CA.

Reason for recall

"Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments."

Distribution

"Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1431-20138mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.697 units - all productsClass IITerminated5 Jun 2013
Z-1432-20138mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.697 - all productsClass IITerminated5 Jun 2013
Z-1433-20138mm Fenestrated Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.697 - all productsClass IITerminated5 Jun 2013

Context

90%of 2013 device events are Class II1,236 events that year
3product records in this eventfirm median: 1
105events on this register for the firm

Other enforcement events for Intuitive Surgical, Inc.

InitiatedProduct (first record)TypeClassRecords
24 Apr 2026Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product DeDeviceClass II1
22 Apr 2026Da Vinci ASSY, PSS, SP1098, Part Number: 380601DeviceClass II1
2 Apr 2026Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200DeviceClass II1
11 Mar 2026Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06DeviceClass I1
30 Dec 2025da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45DeviceClass II1
15 Dec 2025da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,DeviceClass II1
9 Dec 2025Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/DeviceClass II2
9 Dec 2025Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved ScissorDeviceClass II4
4 Dec 2025Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Software Version: N/ADeviceClass II1
17 Nov 2025Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.DeviceClass II1

All 105 events for Intuitive Surgical, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 18 Feb 2014. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.