Device recall: Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion…

Class II FDA enforcement event 64987. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Two sections that are designated for the 8mm and 9mm screws were switched; the holes that hold the 8mm screws are designed for 9mm screws, and the holes that hold the 9mm screws are designated for 8mm screws.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

SpineFrontier, Inc. started this device recall on 5 Oct 2012. The firm address in the FDA record is in Beverly, MA. FDA put the recall in Class II. FDA event 64987 has 1 product record. The first FDA enforcement report date is 7 Aug 2013. The record gives the status Terminated, with the termination date 29 Jul 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US distribution in states of: CO, DC, MA and TX.

Recalling firm

SpineFrontier, Inc., Beverly, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 64987: product records
Recall numberProductQuantityClassStatusReport date
Z-1869-2013Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion within the case, 8mm and 9mm pedicle screws used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients11 caddiesClass IITerminated7 Aug 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
26

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 64987. https://recallregister.com/fda/ii/64987/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US distribution in states of: CO, DC, MA and TX.
What quantity of product is in FDA event 64987?
The FDA record gives this quantity: 11 caddies
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.