Device recall: Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited…

Class II FDA enforcement event 64740. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Ge Healthcare It started this device recall on 4 Mar 2011. The firm address in the FDA record is in Barrington, IL. FDA put the recall in Class II. FDA event 64740 has 1 product record. The first FDA enforcement report date is 22 Jan 2014. The record gives the status Terminated, with the termination date 16 May 2016. 31 of the 114 device events that started in 2011 have more product records. The median device event of 2011 has 1 product record. The openFDA file has 114 device events that started in 2011 and 37 that started in 2010.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.

Recalling firm

Ge Healthcare It, Barrington, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 64740: product records
Recall numberProductQuantityClassStatusReport date
Z-0773-2014Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data…43 unitsClass IITerminated22 Jan 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2011
114
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
17

Other events of the firm

Other FDA enforcement events of Ge Healthcare It
InitiatedProduct (first record)TypeClassRecords
26 Aug 2013Merge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centers to use a siDeviceClass II1
11 Jun 2013The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by tDeviceClass II1
19 Apr 2013Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices DeviceClass II1
12 Mar 2013The Centricity Laboratory System is intended to be an information system designed to support the clinical and administraDeviceClass II1
12 Mar 2013The Centricity Laboratory System is intended to be an information system designed to support the clinical and administraDeviceClass II1
11 Mar 2013Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery syDeviceClass II1
18 Feb 2013Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management sysDeviceClass II1
20 Nov 2012Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information system desigDeviceClass II1
20 Nov 2012The Centricity Laboratory System is intended to be an information system designed to support the clinical and administraDeviceClass II1
20 Nov 2012The Centricity Laboratory System is intended to be an information system designed to support the clinical and administraDeviceClass III1

Ge Healthcare It: all 17 events

Official channels

Cite this page

Recall Register. FDA enforcement event 64740. https://recallregister.com/fda/ii/64740/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.
What quantity of product is in FDA event 64740?
The FDA record gives this quantity: 43 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2011 have the same class?
89% of the 114 device events that started in 2011 are in Class II.