Device recall: ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissue…
Class II FDA enforcement event 64706. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
DePuy Mitek, Inc., a Johnson & Johnson Co. started this device recall on 22 Mar 2013. The firm address in the FDA record is in Raynham, MA. FDA put the recall in Class II. FDA event 64706 has 5 product records. The first FDA enforcement report date is 1 May 2013. The record gives the status Terminated, with the termination date 24 Feb 2014.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Nationwide Distribution.
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co., Raynham, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
5 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1172-2013 | ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. | 65062 | Class II | Terminated | 1 May 2013 |
| Z-1173-2013 | ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. | 84174 | Class II | Terminated | 1 May 2013 |
| Z-1174-2013 | ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. | 206540 | Class II | Terminated | 1 May 2013 |
| Z-1175-2013 | ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. | 46732 | Class II | Terminated | 1 May 2013 |
| Z-1176-2013 | ORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 223116. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. | 46501 | Class II | Terminated | 1 May 2013 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2013
- 1,236
- Share of these events in Class II
- 90%
- Product records in this event
- 5Median of the firm: 2
- Events of the firm on the register
- 16
Other events of the firm
Official channels
- FDA enforcement report Z-1172-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 64706. https://recallregister.com/fda/ii/64706/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Nationwide Distribution.
- What quantity of product is in FDA event 64706?
- The event has 5 product records. The first record gives this quantity:
65062
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2013 have the same class?
- 90% of the 1,236 device events that started in 2013 are in Class II.