Recall Register

Device recall: LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood…

Class II FDA enforcement event 64478. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

When OneTouch Ultra and OneTouch Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Lifescan Inc started this device recall on 25 Feb 2013. The firm address in the FDA record is in Milpitas, CA. FDA put the recall in Class II. FDA event 64478 has 2 product records. The first FDA enforcement report date is 24 Apr 2013. The record gives the status Terminated, with the termination date 23 Oct 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO,…

Recalling firm

Lifescan Inc, Milpitas, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 64478: product records
Recall numberProductQuantityClassStatusReport date
Z-1129-2013LifeScan brand OneTouch Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels.132,301 units totalClass IITerminated24 Apr 2013
Z-1130-2013LifeScan brand OneTouch Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.132,301 units totalClass IITerminated24 Apr 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of Lifescan Inc
InitiatedProduct (first record)TypeClassRecords
11 Mar 2013All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio I Event 64624DeviceClass I1

Official channels

Cite this page

Recall Register. FDA enforcement event 64478. https://recallregister.com/fda/ii/64478/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO,…
What quantity of product is in FDA event 64478?
The event has 2 product records. The first record gives this quantity: 132,301 units total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?