Device recall: qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3…
Class II FDA enforcement event 63504. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Spacelabs Healthcare, Llc started this device recall on 19 Oct 2012. The firm address in the FDA record is in Issaquah, WA. FDA put the recall in Class II. FDA event 63504 has 1 product record. The first FDA enforcement report date is 28 Nov 2012. The record gives the status Terminated, with the termination date 22 May 2013.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as…
Recalling firm
Spacelabs Healthcare, Llc, Issaquah, WA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0403-2013 | qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.1 inches) LCD display with 1024 x 768 resolution. The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or… | 742 total devices distributed; 530 with the U.S. and 212 Internationally. | Class II | Terminated | 28 Nov 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 10
Other events of the firm
Official channels
- FDA enforcement report Z-0403-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 63504. https://recallregister.com/fda/ii/63504/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as…
- What quantity of product is in FDA event 63504?
- The FDA record gives this quantity:
742 total devices distributed; 530 with the U.S. and 212 Internationally.
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?
- 88% of the 823 device events that started in 2012 are in Class II.