Device recall: bioMerieux BacT/ALERT Control Module, catalog number 210147, Clinical instrument The product is most commonly used to…

Class II FDA enforcement event 63467. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The product may mislabel patient information on culture specimen bottles due to a software miscommunication.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Biomerieux Inc started this device recall on 16 Oct 2012. The firm address in the FDA record is in Hazelwood, MO. FDA put the recall in Class II. FDA event 63467 has 31 product records. The first FDA enforcement report date is 28 Nov 2012. The record gives the status Terminated, with the termination date 28 Oct 2013. 2 of the 823 device events that started in 2012 have more product records. The median device event of 2012 has 1 product record. The openFDA file has 823 device events that started in 2012 and 114 that started in 2011.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA including AR, AZ, CA, FL, KS, LA, MD, MN, MT, NC, NV, NY, OR, PA, RI, SC, TX, VA and WA. Internationally to Argentina, Armenia, Austria, Bangladesh, Bosnia and…

Recalling firm

Biomerieux Inc, Hazelwood, MO.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

31 product records in the event.

FDA event 63467: product records
Recall numberProductQuantityClassStatusReport date
Z-0299-2013bioMerieux BacT/ALERT Control Module, catalog number 210147, Clinical instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources…70 modulesClass IITerminated28 Nov 2012
Z-0300-2013bioMerieux BacT/ALERT Combo Module, catalog number, 200290, Clinical instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources…375 modulesClass IITerminated28 Nov 2012
Z-0301-2013bioMerieux BacT/ALERT 3D 60, catalog number, 248009, Clinical instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources…145 modulesClass IITerminated28 Nov 2012
Z-0302-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML01, Clinical instrument software, English, Spanish, Italian, catalog number 411814 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…1336 kitsClass IITerminated28 Nov 2012
Z-0303-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML 02, Clinical instrument software, English, Spanish, Italian, catalog number 411815 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…860 kitsClass IITerminated28 Nov 2012
Z-0304-2013bioMerieux Kit, Instrument & Update, 3D B.40 BTA3D-ZH01, Clinical instrument software, Chinese, catalog number 411816 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms…30 kitsClass IITerminated28 Nov 2012
Z-0305-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML03, Clinical instrument software, Danish, Greek, Polish, Portuguese, catalog number 412623 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with…250 kitsClass IITerminated28 Nov 2012
Z-0306-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML04, Clinical instrument software, Czech, Estonian, Hungarian, Romanian, catalog number 412624 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill…N/AClass IITerminated28 Nov 2012
Z-0307-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML05, Clinical instrument software, Lithuanian, Latvian, Russian, Slovak, catalog number 412625 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill…163 kitsClass IITerminated28 Nov 2012
Z-0308-2013bioMerieux Kit, Instrument & Update, 3D B.40 ZIP-ML06, Clinical instrument software, Japanese, Korean, Norwegian, Brazilian, Portuguese, Turkish, catalog number 412626 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients…671 kitsClass IITerminated28 Nov 2012
Z-0309-2013bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML01, Clinical instrument software, English, Spanish, Italian, catalog number 411818 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…620 kitsClass IITerminated28 Nov 2012
Z-0310-2013bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML02, Clinical instrument software, German, French, Swedish, catalog number 411925 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and…556 kitsClass IITerminated28 Nov 2012
Z-0311-2013bioMerieux Kit, USB HDW, BTA3D60-B-B.40, Clinical instrument software, USB update and mounting hardware-English, catalog number 411956 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…205 kitsClass IITerminated28 Nov 2012
Z-0312-2013bioMerieux Kit, USB HDW, BTA3D COMBO-B-B.40, Clinical instrument software, USB update and mounting hardware-English, catalog number 411957 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with…205 kitsClass IITerminated28 Nov 2012
Z-0313-2013bioMerieux Kit, USB HDW, BTA3D CONTROL-B-B.40, Clinical instrument software, USB update and mounting hardware-English, catalog number 411958 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with…250 kitsClass IITerminated28 Nov 2012
Z-0314-2013bioMerieux Kit, USB Instrument & Update, 3D B.40 USB-ML03, Clinical instrument software, Danish, Greek, Polish, Portuguese, catalog number 412559 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill…29 kitsClass IITerminated28 Nov 2012
Z-0315-2013bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML04, Clinical instrument software, Czech, Estonian, Hungarian, Romanian, catalog number 412620 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill…28 kitsClass IITerminated28 Nov 2012
Z-0316-2013bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML05, Clinical instrument software, Lithuanian, Latvian, Russian, Slovak, catalog number 412621 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill…12 kitsClass IITerminated28 Nov 2012
Z-0317-2013bioMerieux Kit, Instrument & Update, 3D B.40 USB-ML06, Clinical instrument software, Japanese, Korean, Norwegian, Brazilian, Portuguese, Turkish, catalog number 412622 Software update for the clinical instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients…69 kitsClass IITerminated28 Nov 2012
Z-0318-2013bioMerieux BacT/ALERT Control Module, catalog number 210148, Industrial instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of…2 modulesClass IITerminated28 Nov 2012
Z-0319-2013bioMerieux BacT/ALERT Combo Module, catalog number 200291, Industrial instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources…26 ModulesClass IITerminated28 Nov 2012
Z-0320-2013bioMerieux BacT/ALERT Combo Module, catalog number 247014, Industrial instrument The product is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources…2 modulesClass IITerminated28 Nov 2012
Z-0321-2013bioMerieux Kit, Instrument & Update, 3D IND ZIP-ML 01, Industry instrument software update, English, Spanish, Italian, catalog number 412005 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and…288 kitsClass IITerminated28 Nov 2012
Z-0322-2013bioMerieux Kit, Instrument & Update, 3D IND USB-ML 02, Industry instrument software update, English, Spanish, Italian, catalog number 412006 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and…20 kitsClass IITerminated28 Nov 2012
Z-0323-2013bioMerieux Kit, Instrument & Update, B.40 Dualt-ML, Industry instrument software update, English, Spanish, Italian, German, French, Japanese, Brazilian, Portuguese, catalog number 412950 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric…12 kitsClass IITerminated28 Nov 2012
Z-0324-2013bioMerieux Kit, Instrument & Update, B.40, 3D IND USB-ML01, Industry instrument software update, English, Spanish, Italian, catalog number 412002 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…83 kitsClass IITerminated28 Nov 2012
Z-0325-2013bioMerieux Kit, Instrument & Update, B.40, 3D IND USB-ML02, Industry instrument software update, German, French, catalog number 412003 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and…22 kitsClass IITerminated28 Nov 2012
Z-0326-2013bioMerieux Kit, Instrument & Update, B.40, 3D I USB-ML01, Industry instrument software update, English, Spanish, Italian, catalog number 413481 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…23 kitsClass IITerminated28 Nov 2012
Z-0327-2013bioMerieux Kit, Instrument & Update, B.40, 3D I USB-ML02, Industry instrument software update, German, French, catalog number 413482 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms…10 kitsClass IITerminated28 Nov 2012
Z-0328-2013bioMerieux Kit, Instrument & Update, B.40, 3D I ZIP-ML01, Industry instrument software update, English, Spanish, Italian, catalog number 413483 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…33 kitsClass IITerminated28 Nov 2012
Z-0329-2013bioMerieux Kit, Instrument & Update, B.40, 3D I ZIP-ML02, Industry instrument software update, English, Spanish, Italian, catalog number 413484 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs…5 kitsClass IITerminated28 Nov 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
31Median of the firm: 1
Events of the firm on the register
58

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 63467. https://recallregister.com/fda/ii/63467/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA including AR, AZ, CA, FL, KS, LA, MD, MN, MT, NC, NV, NY, OR, PA, RI, SC, TX, VA and WA. Internationally to Argentina, Armenia, Austria, Bangladesh, Bosnia and…
What quantity of product is in FDA event 63467?
The event has 31 product records. The first record gives this quantity: 70 modules
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.