Device recall: Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part…

Class II FDA enforcement event 63204. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Orthovita, Inc., dBA Stryker Orthobiologics. started this device recall on 10 Nov 2012. The firm address in the FDA record is in Malvern, PA. FDA put the recall in Class II. FDA event 63204 has 13 product records. The first FDA enforcement report date is 5 Jun 2013. The record gives the status Terminated, with the termination date 29 Oct 2014.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.

Recalling firm

Orthovita, Inc., dBA Stryker Orthobiologics., Malvern, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

13 product records in the event.

FDA event 63204: product records
Recall numberProductQuantityClassStatusReport date
Z-1387-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0502 Product Usage: Facilitate placement of bone cement in weakened or diminished bone78Class IITerminated5 Jun 2013
Z-1388-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 6", Part Number 2090-0505 Product Usage: Facilitate placement of bone cement in weakened or diminished bone164Class IITerminated5 Jun 2013
Z-1389-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0501 Product Usage: Facilitate placement of bone cement in weakened or diminished bone411Class IITerminated5 Jun 2013
Z-1390-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Syringe Delivery System, 6", Part Number 2090-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone155Class IITerminated5 Jun 2013
Z-1391-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 4", Part Number 2110-0505 Product Usage: Facilitate placement of bone cement in weakened or diminished bone35Class IITerminated5 Jun 2013
Z-1392-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 8", Part Number 2110-0526 Product Usage: Facilitate placement of bone cement in weakened or diminished bone27Class IITerminated5 Jun 2013
Z-1393-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger, 6", Part Number 2110-0501 Product Usage: Facilitate placement of bone cement in weakened or diminished bone414Class IITerminated5 Jun 2013
Z-1394-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Diamond Needle, 4", Part Number 2110-0506 Product Usage: Facilitate placement of bone cement in weakened or diminished bone19Class IITerminated5 Jun 2013
Z-1395-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 8", Part Number 2110-0525 Product Usage: Facilitate placement of bone cement in weakened or diminished bone5Class IITerminated5 Jun 2013
Z-1396-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 6", Part Number 2110-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone7Class IITerminated5 Jun 2013
Z-1397-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger, 8", Part Number 2110-0502 Product Usage: Facilitate placement of bone cement in weakened or diminished bone17Class IITerminated5 Jun 2013
Z-1398-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 6", Part Number 2110-0524 Product Usage: Facilitate placement of bone cement in weakened or diminished bone12Class IITerminated5 Jun 2013
Z-1399-2013Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Catheter, 6", Part Number 2110-0503 Product Usage: Facilitate placement of bone cement in weakened or diminished bone69Class IITerminated5 Jun 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
13Median of the firm: 2
Events of the firm on the register
4

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 63204. https://recallregister.com/fda/ii/63204/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
What quantity of product is in FDA event 63204?
The event has 13 product records. The first record gives this quantity: 78
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.