Device recall: GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused…
Class II FDA enforcement event 62729. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
GE OEC Medical Systems, Inc started this device recall on 26 Apr 2010. The firm address in the FDA record is in Salt Lake City, UT. FDA put the recall in Class II. FDA event 62729 has 1 product record. The first FDA enforcement report date is 29 Aug 2012. The record gives the status Terminated, with the termination date 22 Aug 2012.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide and Nationwide (USA) distribution.
Recalling firm
GE OEC Medical Systems, Inc, Salt Lake City, UT.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2241-2012 | GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5.… | 845 units | Class II | Terminated | 29 Aug 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2010
- 37
- Share of these events in Class II
- 89%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 28
Other events of the firm
Official channels
- FDA enforcement report Z-2241-2012The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 62729. https://recallregister.com/fda/ii/62729/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide and Nationwide (USA) distribution.
- What quantity of product is in FDA event 62729?
- The FDA record gives this quantity:
845 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2010 have the same class?
- 89% of the 37 device events that started in 2010 are in Class II.