Device recall: Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4…
Class II FDA enforcement event 62540. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Zimmer, Inc. started this device recall on 18 Jul 2012. The firm address in the FDA record is in Warsaw, IN. FDA put the recall in Class II. FDA event 62540 has 4 product records. The first FDA enforcement report date is 22 Aug 2012. The record gives the status Terminated, with the termination date 17 Apr 2014.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
Recalling firm
Zimmer, Inc., Warsaw, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
4 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2212-2012 | Trabecular Metal Reverse Shoulder Liner 40mm, part 00-4309-028-00 - Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly… | 185 (162 in USA) | Class II | Terminated | 22 Aug 2012 |
| Z-2213-2012 | 00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 40mm 65¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the… | 252 (210 in USA) | Class II | Terminated | 22 Aug 2012 |
| Z-2214-2012 | 00-4309-029-00- Trabecular Metal Reverse Shoulder Liner 36mm, Nonsterile Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly… | 75 (51 in USA) | Class II | Terminated | 22 Aug 2012 |
| Z-2215-2012 | 00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on… | 212 (179 USA) | Class II | Terminated | 22 Aug 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class II
- 88%
- Product records in this event
- 4Median of the firm: 1
- Events of the firm on the register
- 108
Other events of the firm
Official channels
- FDA enforcement report Z-2212-2012The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 62540. https://recallregister.com/fda/ii/62540/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
- What quantity of product is in FDA event 62540?
- The event has 4 product records. The first record gives this quantity:
185 (162 in USA)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?
- 88% of the 823 device events that started in 2012 are in Class II.