Recall Register

Device recall: ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial…

Class II FDA enforcement event 62456. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Stryker Howmedica Osteonics Corp. started this device recall on 30 Apr 2012. The firm address in the FDA record is in Mahwah, NJ. FDA put the recall in Class II. FDA event 62456 has 2 product records. The first FDA enforcement report date is 1 Aug 2012. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Distribution

Recalling firm

Stryker Howmedica Osteonics Corp., Mahwah, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 62456: product records
Recall numberProductQuantityClassStatusReport date
Z-2089-2012ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial Estate, Carrigtwohill County Cork, Ireland. The ABG II Modular range of implants is constituted of 8 right stems and 8 left stems. The necks are made of GADS Vitallium, a proprietary Co Cr alloy that…9003 unitsClass IIOngoing1 Aug 2012
Z-2090-2012Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy,…(NLS/NLV) - 23193 units and (SPT) - 20,743 unitsClass IIOngoing1 Aug 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
59

Other events of the firm

Other FDA enforcement events of Stryker Howmedica Osteonics Corp.
InitiatedProduct (first record)TypeClassRecords
26 Apr 2017Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 ProducDeviceClass II1
2 Dec 2016Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an instrument DeviceClass II1
9 Nov 2016GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-DeviceClass II1
29 Aug 2016LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, 36 mm, REF 6260-9-236; Modular components designed DeviceClass II1
18 Aug 2016Stryker Orthopaedics Universal Acetabular Cup Positioner/Impactor, REF 2101-0200, Non-Sterile, Howmedica Osteonics Corp.DeviceClass II1
1 Aug 2016Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio PDeviceClass II1
1 Jul 2016Orthopedic manual surgical instrument Usage: Exeter Rasp Handle is a reusable instrument which is part of the Exeter femDeviceClass II1
30 Jun 2016Supracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stable internal fixatDeviceClass II2
23 Jun 2016Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S ProdDeviceClass II1
23 Jun 2016Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement orDeviceClass II1

Stryker Howmedica Osteonics Corp.: all 59 events

Official channels

Cite this page

Recall Register. FDA enforcement event 62456. https://recallregister.com/fda/ii/62456/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Distribution
What quantity of product is in FDA event 62456?
The event has 2 product records. The first record gives this quantity: 9003 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.